4.2
Patient Information (Optional)
The user may enter a patient identification, such as the patient’s initials or medical record
number, into the SAPS if the user chooses. If the user does not provide a patient identification,
the IPG’s unique serial number can be used.
1. Enter the patient identification into the
Patient ID
field, as shown in Figure 4.10.
2. Program this Patient ID to the IPG by selecting the
Download Settings to IPG
button,
located in the center of the screen (see Figure 4.1).
This is only required to be performed once. SAPS will recognize the programmed Patient ID
each time the IPG is interrogated. The IPG retains this information indefinitely.
Figure 4.10: Patient ID
4.3
Implant Settings
To enable each of the therapy modes, the button next to the desired
Therapy Mode
is selected
and turned on.
1. Turn on your desired
Therapy Mode
to adjust the
Implant Settings
.
2. Each of the implant setting parameters available in the
Implant Settings
menu has a box
next to the setting (see Figure 4.11). Select the box to display all available parameters in a
drop-down box.
3. Select the appropriate value for each parameter, and program the parameters into the IPG
using the
Download Settings to IPG
function. Examples of parameters are shown in
Figure 4.12 through Figure 4.15.
4. Once you choose new implant settings, always verify that the implant setting parameters
chosen are tolerable for the patient using the
Stim Test Mode
function. See
Section
Once the values programmed via the
Test
function or the
Download Settings to IPG
function
are input into the IPG, all subsequent therapy sessions will be performed at these parameters.
The user can modify the parameters at any time by selecting new parameters and programming
them into the IPG.
Summary of Contents for 2100
Page 22: ...26 0002 0005 0 en 2022 22 Figure 4 13 Frequency Figure 4 14 Train Duration ...
Page 35: ...26 0002 0005 0 en 2022 35 Figure 5 10 Magnet Swipe ...
Page 44: ...26 0002 0005 0 en 2022 44 Figure 8 2 Patient Log File Export Log ...
Page 45: ...26 0002 0005 0 en 2022 45 9 INTENTIONALLY BLANK ...