LVS-95XX Series Barcode Quality Station Operations Manual
LVS-95XX Series Barcode Quality Station Operations Manual - English
Page 94 of 94
Window Scratches:
When the sensor window is scratched, it may cause a streak in the image. This streak may cause the
barcode grade results to be distorted if the streak is located within the barcode image. When a
scratch occurs, you may need to return the system to Microscan (or your Microscan distributor) for
sensor replacement. Contact your Microscan representative for more information.
Replace the Calibrated Conformance Standard Test Card every two years.
LVS-9580
Cleaning Instructions:
The LVS-9580 window may need to be cleaned daily, depending on use. Debris on the window may
cause the LVS-9580 to not grade accurately.
Locate the following supplies:
•
Commercially available household glass cleaner, such as Windex®, Glassex®, VISS®, or
Mr. Muscle®. DO NOT use an industrial-strength glass cleaner.
•
Soft, lint-free, non-abrasive towel or cloth
Dampen the cloth with the household glass cleaner and gently wipe the window. Inspect the window
closely, looking for any label debris that may be stuck on the window. Do not scrape the window with
a sharp object as this may damage the window. Any damage to the window will be detected during
the calibration process.
•
DO NOT directly spray the window with glass cleaner; always spray a towel or cloth with
household glass cleaner and then gently wipe the window.
•
DO NOT use an industrial-strength glass cleaner.
Replace the Calibrated Conformance Standard Test Card every two years.
LVS-9505
Replace the Calibrated Conformance Standard Test Card every two years.
The LVS-9505 light tray has LED lights which have a life span of approximately 10,000 hours; thus,
there should be no replacement concerns with the light tray. Contact Microscan or your Microscan
distributor with any questions.
LVS-9500
Monthly:
Using 60 to 80 PSI (400-480 kPa) of positive air flow, clean the lens mounted inside the LVS-
9500 housing; this ensures the removal of small particles on the surface of the lens.
Vacuum the two air intake holes located under the main housing; this removes any build-up of
particles trapped in the two filters.
For Systems Using a DC Fluorescent Light Tray:
Replace the DC Fluorescent Light Tray every
two years; these special lights diminish in brightness over several years. Replacing the Light Tray
every two years ensures continuous calibration levels.
For Systems Using LED Lights:
The LVS-9500 light tray has LED lights which have a life span of
approximately 10,000 hours; thus, there should be no replacement concerns with the light tray.
Contact Microscan or your Microscan distributor with any questions.
Replace the Calibrated Conformance Standard Test Card every two years.