11
Automated triple measurement
In
«1st VISIT» Mode, the WatchBP Office device
automatically takes three consecutive measurements
at one minute* intervals by default. The result of
these three measurements are averaged to produce
the first visit blood pressure measurement.
Determine proper measurement arm
If the measurement result between the two arms
differs more than 20mmHg for Systolic or 10mmHg
for Diastolic for all three measurements, the device
will indicate the higher measurement arm by a
flashing L or R.
Three operation modes (cont.)
1
2
3
60 sec
.
60
sec.
*
The user can manually set up the measurement
interval of 15, 30, 45 or 60 seconds in 1st Visit and
Follow-up Mode. (Please refer to special functions
section page.24 “Set up measurement interval”).
The higher arm should be used as the preferred
measurement arm for future measurements.