6
This device complies with the requirements of the Medical Device
Directive 93/42/EEC.
Technical alterations reserved.
According to the Medical Product User Act a biennial technical
inspection is recommended for professional users. Please
observe the applicable disposal regulations.
14.www.microlife.com
Detailed user information about our thermometers and blood pres-
sure monitors as well as services can be found at www.micro-
life.com.
Reference body site:
Sublingual
Measurement range:
Body mode: 32.0 - 43.0 °C / 89.6 - 109.4
°F
Resolution:
0.1 °C / °F
Measurement accu-
racy (Laboratory):
Body mode:
±0.2 °C, 35.0 ~ 42.0 °C / ±0.4 °F, 95.0
~
107.6 °F
±0.3 °C, 32.0
~
34.9 °C and 42.1
~
43.0
°C /
±0.5 °F, 89.6
~
94.8 °F and 107.8
~
109.4 °F
Clinical results:
Repeatability: 0.19 °C
Bias: 0.03 °C
Limits of agreement: 1.33 °C
Display:
L
iquid
C
rystal
D
isplay, 4 digits plus
special icons
Acoustic:
The unit is turned ON and ready for the
measurement: 1 short beep.
Complete the measurement: 1 long beep.
System error or malfunction: 3 short
beeps.
Fever alarm: 10 short beeps.
Memory:
30 readings recall in the memory mode
with a record of both time and date.
Backlight:
The display light will be GREEN for
1 second, when the unit is turned ON.
The display light will be GREEN for 5
seconds, when a measurement is
completed with a reading less than
37.5 °C / 99.5 °F.
The display light will be RED for
5 seconds, when a measurement is
completed with a reading equal to or
higher than 37.5 °C / 99.5 °F.
Operating condi-
tions:
10 - 40 °C / 50.0 - 104 °F
15 - 95 % relative maximum humidity
Storage conditions:
-25 - +55 °C / -13 - +131 °F
15 - 95 % relative maximum humidity
Automatic Switch-
off:
Approx. 1 minute after last measurement
has been taken.
Battery:
2 x 1.5 V alkaline batteries; size AAA
Battery lifetime:
approx. 800 measurements
(using new batteries)
Dimensions:
151 x 46 x 60.5 mm
Weight:
60 g (with battery), 58 g (w/o battery)
IP Class:
IP22
Reference to
standards:
ISO 80601-2-56; IEC 60601-1; IEC
60601-1-2 (EMC); IEC 60601-1-11
Expected service life:
5 years or 12000 measurements
Summary of Contents for IR310
Page 52: ...50 1 ACCUsens good 30 10 37 5 C 2 Microlife 15 3 3 m...
Page 64: ...62...