7. Error messages/troubleshooting

If an error occurs during a measurement, the measure-
ment is discontinued and a corresponding error code is
displayed (example: Error no. 2).

Error No. Possible cause(s)/Solutions

ERR 1

The tube may have loosened, or no pulse was detect-
ed.* Ensure cuff connections are tight with proper cuff
placement. See section 5.3.

ERR 2

Unnatural pressure impulses influenced the measure-
ment result. Reason: The arm was moved during the
measurement (artefact). Repeat measurement,
keeping still and quiet.

ERR 3

Inflation of the cuff takes too long. The cuff is not
correctly seated or the hose connection is not tight.
Re-position cuff and repeat the measurement.

ERR 5

The measured readings indicated an unacceptable
difference between systolic and diastolic pressures.
Take another reading following directions carefully.
Contact your doctor if you continue to get unusual
readings.

HI

The cuff pressure is too high. Relax for 5 minutes
and repeat the measurement.*

LO

The pulse is too low (less than 40). Repeat the
measurement.*

*If this or any other problem occurs repeatedly, please consult
your doctor.

19

Summary of Contents for bp3gx1-5x

Page 1: ......

Page 2: ...batteries 4 2 Setting the date and time 4 3 Using the AC power adapter 4 4 Cuff tube connection 4 5 Select the user 4 6 Select the measuring mode standard or Microlife Averaging Mode 4 7 Microlife Averaging Mode 5 Measurement procedure 5 1 Before measurement 5 2 Common sources of error 5 3 Fitting the cuff 5 4 Measuring procedure 5 5 Memory displaying the last 99 measurements 5 6 Discontinuing a m...

Page 3: ...not suitable for checking the frequency of heart pacemakers In cases of irregular heartbeat measurements made with this instru ment should only be evaluated after consultation with your doctor Electromagnetic interference The device contains sensitive electronic components Avoid strong electri cal or electromagnetic fields in the direct vicinity of the device e g mobile telephones microwave ovens ...

Page 4: ... values that are too low i e systolic values under 105 mmHg and or diastolic values under 60 mmHg consult your physician Even with normal blood pressure values a regular self check with your blood pressure monitor is recommended You can detect possible changes in your values early and react appropriately If you are undergoing medical treatment to control your blood pressure keep a record of values...

Page 5: ... supply of blood to important organs heart brain muscles can result from arteriosclerosis Furthermore the heart will become structurally damaged with increased blood pressure values c There are many different causes of high blood pressure We differenti ate between the common primary essential hypertension and second ary hypertension The latter group can be ascribed to specific organ malfunctions P...

Page 6: ...ical examination do regular exercise Choose sports which require stamina and avoid those which require strength Avoid reaching the limit of your performance With previous illnesses and or an age of over 40 years please consult your doctor before beginning your exercise routine You must receive advice regarding the type and extent of exercise that is appropriate for you ...

Page 7: ...hould be measured with a measuring tape in the middle of the relaxed upper arm Do not force cuff connection into the opening Make sure the cuff connection is not pushed into the AC adapter port If the cuff is too small call 1 800 568 4147 for further information Cuff Socket yellow Time Date Button POWER Button Mode Switch Select 1 or 3 Measurements USB Port Memory Recall Button AC Adapter Port Hyp...

Page 8: ...device won t function until the batter ies have been replaced Please use AA Long Life or Alkaline 1 5 V batteries If the blood pressure monitor is not used for long periods remove the batteries from the device 4 2 Setting the date and time 1 After the new batteries are activated the year number flashes in the display You can advance the year by pressing the M button To confirm and then set the mon...

Page 9: ...er and cable are not damaged b Plug the adapter cable into the AC adapter port on the right side of monitor c Plug the AC adapter into a 110 V power socket U S or Canada d Test that power is available by pressing the POWER button Note No power is taken from the batteries while the AC adapter is connected to the instrument If the power is interrupted during a measurement e g by removal of the adapt...

Page 10: ...ging mode push the switch downward to Position 3 Note the MAM icon will illuminate on the display If you select 1 then only one measurement will be taken If you select 3 the unit will inflate and deflate three times resulting in one final average 4 7 Microlife Averaging Mode MAM In Microlife Averaging Mode MAM three measurements are automati cally taken in succession and the result is then automat...

Page 11: ...able relaxed position and do not flex any of the muscles in the measurement arm during the measurement Use a cushion for support if necessary If the arm artery lies considerably lower or higher than the heart an erroneously high or low blood pressure will be measured Each 15 cm 6 difference in height between your heart and the cuff results in a measurement error of 10 mmHg Cuffs that are too narro...

Page 12: ...re the cuff wrap it around your arm and press the hook and loop material together Check the size range indicator on the cuff The green arrow should point within the green size range on the inside of the cuff d There should be little free space between the arm and the cuff You should be able to fit 2 fingers between your arm and the cuff Clothing must not restrict the arm Any piece of clothing whic...

Page 13: ...reasing cuff pressure is continually shown b After automatically reaching an individual pressure the pump stops and the pressure slowly falls The cuff pressure is displayed during the measurement c When the device has detected your pulse the heart symbol in the display begins to blink d When the measurement has been concluded the air will automatically release from the cuff The measured systolic a...

Page 14: ... as a single averaged measurement individual results are not displayed There is a 15 second resting time between each measurement A count down indicates the remaining time between measurements If one of the measurements causes an error message it will be repeated one more time If any additional error occurs the measurement will be discontinued and error code displayed 13 ...

Page 15: ...n to scroll to that stored reading Memory full When the memory has stored 99 results a new measured value is stored by overwriting the oldest value Clear all values If you are sure that you want to permanently remove all stored values hold down the M button the instrument must have been switched off beforehand until CL appears and then release the button If you do not want to clear the values pres...

Page 16: ...al for details of the classifications A triangle is displayed to the right of the traffic light bar according to your measurement If your triangle is in the green zone your measurement is Normal according to NIH standards If your triangle is in the yellow zone your measurement is Pre Hypertension If your triangle is in the orange zone it is Stage 1 Hypertension If your triangle is in the dark oran...

Page 17: ... show your doctor the following explanation Information on frequent appearance of the irregular heartbeat symbol This instrument is an oscillometric blood pressure monitor device that also analyzes pulse frequency during measurement The instrument is clinically tested If pulse irregularities occur during the measurement the irregular heartbeat symbol is displayed with the measurement If the symbol...

Page 18: ...d time and possibly also set alarm times must be reset the year number will flash automati cally after the batteries are replaced 4 To set date and time follow the procedure described in Section 4 3 Note Use four new Long Life 1 5 V AA batteries Do not use batteries beyond their expiration date If the monitor is not going to be used for a prolonged period the batteries should be removed Using rech...

Page 19: ...on and data transmission a Insert CD into CD ROM drive of your PC The installation will start automatically If not click on SETUP EXE b Connect the monitor via USB cable with the PC Three horizontal bars will appear on the display and last for 3 seconds c The bars will then flash to indicate that the connection between computer and device is successfully made As long as the cable is plugged the ba...

Page 20: ...surement artefact Repeat measurement keeping still and quiet ERR 3 Inflation of the cuff takes too long The cuff is not correctly seated or the hose connection is not tight Re position cuff and repeat the measurement ERR 5 The measured readings indicated an unacceptable difference between systolic and diastolic pressures Take another reading following directions carefully Contact your doctor if yo...

Page 21: ...easure blood pressure values or the values measured are too low or high 20 Please read the following informa tion and points listed in Section 5 2 Common sources of error Repeat the measurement Please note Blood pressure fluctuates continually so succes sive measurements will show some variability Record the daily development of the measured values and consult your doctor Please note Individuals v...

Page 22: ...low the tubing to kink and keep it away from sharp edges e Do not drop the monitor or treat it roughly in any way Avoid strong vibrations f Never open the monitor This invalidates the manufacturer s warranty g Batteries and electronic instruments must be disposed of in accordance with the locally applicable regulations not with domestic waste Further information Blood pressure is subject to fluctu...

Page 23: ...nal standards your monitor should be checked for accuracy every 2 years 10 Certifications Device standard Device corresponds to the requirements of the standard for non invasive blood pressure monitors AAMI ANSI SP10 IEC 60601 1 IEC 60601 1 2 Electromagnetic compatibility Device fulfills the stipula tions of the International standard IEC 60601 1 2 Clinical testing Clinical performance tests were ...

Page 24: ...range 0 299 mmHg Memory Automatically stores the last 99 measurements for 2 users total 198 Measuring resolution 1 mmHg Accuracy Pressure within 3 mmHg or 2 of reading 200 mmHg Pulse 5 of the reading Power source a 4 AA batteries 1 5 V b AC adapter 6 V DC 600 mA voltage 4 5 V DC to 6 V DC Accessories Cuff type Wide range cuff for arm circumference 22 42 cm 8 7 16 5 BP analyzer software CD USB A to...

Page 25: ... How to contact us Microlife USA Inc 1617 Gulf to Bay Blvd 2nd Floor Clearwater FL 33755 Toll Free Help Line 1 800 568 4147 Email custserv microlifeusa com Fax 727 451 0492 www microlifeusa com Made in China ...

Page 26: ...25 Monitor de Presión Arterial Automatico Manual de Instrucción Modelo BPGX1 5X ...

Page 27: ...terías 4 2 Ajuste de la fecha y tiempo 4 3 Utilización del adaptador de corriente alterna CA 4 4 Conexión del brazalete 4 5 Selecciónar al usuario 4 6 Selección a el modo de medición estándar o modo promedio Microlife 4 7 Technología modo promedio Microlife 5 Procedimiento para la medición 5 1 Antes de la medida 5 2 Fuentes comunes de error 5 3 Ajuste del brazalete 5 4 Procedimiento de medición 5 ...

Page 28: ...27 9 Garantía 10 Certificaciones 11 Especificaciones técnicas 12 Como contactarnos ...

Page 29: ...cia debe usted alterar la dosis de ninguno de los medicamentos prescriptos por su doctor La lectura del pulso no es la indicada para chequear la frecuencia de latidos de los marcapasos En casos de latidos irregulares arritmia las mediciones tomadas con este instrumento deben ser evaluadas sólamente luego de una consulta con su médico Interferencia Electromagnética Este dispositivo contiene compone...

Page 30: ...ea diastólica está entre 90 mmHg y 99 mmHg consulte a su médico Un auto examen periódico es necesario Si los valores de su presión sanguínea son demasiado bajos ej Valores sistólicos por debajo del 150 mmHg y o valores diastólicos por debajo de 60 mmHg consulte con su médico Aún con valores normales de presión se recomienda un auto control regular con su monitor de presión sanguínea Usted podría d...

Page 31: ...e a los órganos importantes corazón cerebro músculos puede resultar de la arteriosclerosis Además el corazón sufriría daños con el aumento de la presión sanguínea c Hay diferentes causas de presión alta Diferenciamos entre común primaria esencial hipertensión y secundaria hipertensión Esta última se debe al mal funcionamiento de los órganos específicos Por favor consulte con su médico para obtener...

Page 32: ...de haber sufrido enfermedades con anterioridad y o tener una edad superior a los 40 años consulte con su médico antes de comenzar a practicar deporte Él le aconsejará sobre el tipo de deporte más apropiado para usted 3 Componentes de su monitor de presión arterial a Unidad de medida Conección del Brazalete amarillo Botón de hora y fecha Botón de Encendido Apagado Selector de modali dad Individual ...

Page 33: ...de presión arterial por primera vez 4 1 Inserción de baterías Una vez desempaquetado el aparato inserte primero las baterías El compartimento de la batería está situado en la parte inferior del aparato a Remueva la cubierta de las baterías b Inserte las baterías 4 AA 1 5 V observando la indicada polaridad c Si el símbolo de la batería aparece en la pantalla significa que las baterías están descarg...

Page 34: ...uego fijar el día presione el botón de la HORA 3 Por favor siga las instrucciones anteriores para configurar el día hora y minutos 4 Cuando usted haya establecido el último minuto y presionado el botón de la HORA la fecha y hora se ajustan y se observarán en la pantalla 5 Si usted desea cambiar la fecha y hora presione y mantenga presionado el botón de la HORA por aproximadamente 3 segundos hasta ...

Page 35: ...otencia US o Canadá d Prueba que la alimentación está disponible pulsando el botón de Encendio Apagado Nota Las baterías no se consumen cuando el adaptador esta conectado a la unidad Si la alimentación se interrumpe durante una medición ej por elimina ción del adaptador de la toma pared el instrumento debe ser desacti vado por la eliminación de la clavija del instrumento Si usted tiene alguna preg...

Page 36: ...sición 1 b Para seleccionar el modo promedio Microlife mueva el botón hacia abajo a la posición 3 Nota el MAM icono se iluminará en la pantalla Si Ud selecciona 1 entonces sólo habrá una medición Si selecciona 3 el monitor se infle y desinfle tres veces Esto resulta en un promedio final 4 7 Technologia modo promedio Microlife MAM Cuando se utiliza la modalidad modo promedio Microlife MAM se tomará...

Page 37: ...ntre en una cómoda y relajada posición y no flexione ninguno de los músculos del brazo durante la medición Utilice un almohadón para soporte si es necesario Si la vena del brazo se encuentra a una posición considerablemente mas baja o alta que el corazón se podría obtener una lectura errónea o muy alta o muy baja Cada 15 cm 6 de diferencia en altura entre el corazón y la bocamanga puede resultar e...

Page 38: ... dor de su brazo y presione el material de gancho y bucle juntos Verifique el indicador de rango de tamaño en el brazalete La flecha verde debe apuntar dentro del rango de tamaño verde en el interior del brazalete d Debe de quedar un poco de espacio entre el brazo y la bocamanga Debería caber 2 dedos entre la bocamanga y su brazo La ropa no debe restringir el brazo Cualquier atuendo que moleste o ...

Page 39: ...se podrá observar continuamente el aumento de la presión b Cuando se llega a una presión individual la bomba se detiene y la presión comienza a bajar lentamente La presión de la bocamanga se despliega en la pantalla durante la medición c Cuando el dispositivo detecta su pulso el símbolo del corazón comienza a titilar en la pantalla por cada latido d Cuando la medición ha concluido el aire se liber...

Page 40: ...ostración de las últimas 99 mediciones Al final de una medición este monitor almacena automáticamente cada resultado con la fecha y la hora Esta unidad guarda 99 memorias para cada uno de los dos usuarios Visualización de los valores almacenados Con la unidad apagada presione el botón M La pantalla le enseñará primero A luego muestrá el promedio de todas las mediciones almace nadas en la unidad No...

Page 41: ... la medición de la presión arterial por cualquier motivo ej el paciente no se siente bien el botón encendio apagado botón puede ser presionado en cualquier momento El dispositivo inmedia tamente comenzará a disminuir automáticamente la presión en el brazale te y entra en modo de reposo 5 7 Indicador de riesgo de hipertensión Las barras en el lado izquierdo de la pantalla muestran el alcance de la ...

Page 42: ...a amarillo su medida es Pre Hipertensión Si su medida tiene una barra naranja claro su medida es Hipertensión en Etapa 1 Si su medida tiene una barra naranja oscuro su medidia es Hipertensión en Etapa 2 41 Indicación de una presión Normal Indicación de una presión Pre Hipertensión Indicación de una presión Hypertensión Etapa 1 Indicación de una presión Hypertensión Etapa 2 ...

Page 43: ...iente explicación Información sobre la frecuente aparición del símbolo de latido irregular Este instrumento es un monitor oscilométrico de presión sanguínea que también analiza la frecuencia del pulso durante la medición El instrumento está clínicamente analizado Si durante la medición ocurren irregularidades del pulso se observará el símbolo de latido irregular Si el símbolo aparece con mas frecu...

Page 44: ...ra posible mente las alarmas deberán ser nuevamente fijados Los números que indican el año centellearán automáticamente luego de que las baterías sean reemplazadas 4 Para fijar la fecha y la hora siga el procedimiento descrito en 4 3 Nota Utilice 4 baterías nuevas de Larga Vida 1 5 V AA No utilice baterías que han expirado Si el dispositivo no será utilizado por un largo periodo es aconsejable que...

Page 45: ... Inserte el CD en el compartimiento correspondien te de su computadora La instalación comenzará automáticamente De lo contrario presione el ícono SETUP EXE b Conecte el monitor a su computador por medio del cable USB con el PC Tres barras horizontales aparecerán en la pantalla por una duración de 3 segundos c Las barras entonces comenzarán a parpadear indicando que la conexión entre ambos disposit...

Page 46: ...ción artefacto Vuelva a repetir la medición manteniendo quieto y en silencio ERR 3 El inflado de la bocamanga toma mucho tiempo La bocamanga no está colocado correctamente o la conexión del tubo no esta bien ajustada Vuelva a colocar el brazalete y repita la medición ERR 5 La diferencia entre lecturas sistólica y diastólica es excesiva Mediase otra vez siguiendo las instruccio nes cuidadosamente C...

Page 47: ... o nada en absoluto Se hace imposible obtener una medición 46 Solución 1 Revise la polaridad de las baterías 2 Si el despliegue es inusual remueva las baterías y cámbielas por nuevas Verifique la conexión del tubo de la bocamanga y conéctela correctamente 1 Ajuste la bocamanga al brazo correctamente 2 Mida la presión otra vez en paz y tranquilidad siguiendo cuidadosamente los detalles en las secci...

Page 48: ...bo cuidadosamente No lo jale No permita que el tubo se doble y manténgalo lejos de los bordes afilados e No deje caer el monitor ni permita maltra tos Evite vibraciones fuertes f Nunca abra el monitor Ello anularía la garantía del fabricante g Deben deshacerse de las baterías y disposi tivo conforme a las regulaciones locales Más información La presión arterial está sujeto a fluctuaciones incluso ...

Page 49: ... no estan cubiertos por la garantía Nota Conforme a las reglas internacionales su monitor debe ser revisado para calibración cada dos años 10 Certificaciones Regulación del dispositivo Este dispositivo para procedi mientos no Invasivos está diseñado conforme a las reglas de fabricación Europeas y de los Estados Unidos de América AAMI ANSI SP10 IEC 60601 1 IEC 60601 1 2 Compatibilidad electromagnét...

Page 50: ...tomáticamente las últimas 99 mediciones para dos usuarios total 198 Medida de resolución 1 mmHg Precisión Presión dentro de un pulso de 3 mmHg o 2 de la lectura 200 mmHg Pulso 5 de lectura Fuente de energía a 4 baterías AA 1 5 V b Adaptador CA 6 V DC 600 mA voltaje 4 5 V DC a 6 V DC Accesorios Bocamanga tipo fácil ajustar para brazo de circunferencia 22 42 cm 8 7 16 5 Programa de análisis de la pr...

Page 51: ...os Microlife USA Inc 1617 Gulf to Bay Blvd 2nd Floor Clearwater FL 33755 Línea de servicio al cliente llame sin cargo 1 800 568 4147 Email custserv microlifeusa com Fax 727 451 0492 www microlifeusa com Hecho en China 50 ...

Page 52: ...51 ...

Page 53: ...BP3GX1 5X 0713 1 ...

Reviews: