Which batteries and which procedure?
Using rechargeable batteries
You can also operate this instrument using rechargeable batteries.
8. Using a Mains Adapter
You can operate this instrument using the Microlife mains adapter
(DC 6V, 600mA).
When the mains adapter is connected, no battery current is
consumed.
9. Error Messages
If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g. «
ERR 3
», is displayed.
* Please consult your doctor, if this or any other problem occurs
repeatedly.
2. Replace the batteries – ensure correct polarity as shown by the
symbols in the compartment.
)
Use 4 new, long-life 1.5V, size AA batteries.
)
Do not use batteries beyond their date of expiry.
)
Remove batteries if the instrument is not going to be used
for a prolonged period.
)
Only use «NiMH» type reusable batteries.
)
Batteries must be removed and recharged when the flat
battery symbol appears. They should not remain inside the
instrument as they may become damaged (total discharge
as a result of low use of the instrument, even when
switched off).
)
Always remove the rechargeable batteries if you do not
intend to use the instrument for a week or more.
)
Batteries cannot be charged in the blood pressure monitor.
Recharge batteries in an external charger and observe the
information regarding charging, care and durability.
)
Only use the Microlife mains adapter available as an orig-
inal accessory appropriate for your supply voltage, e.g. the
«Microlife 230V adapter».
)
Ensure that neither the mains adapter nor the cable are
damaged.
1. Plug the adapter cable into the mains adapter socket
5
in the
blood pressure monitor.
2. Plug the adapter plug into the wall socket.
Error
Description Potential cause and remedy
«
ERR 1
» Signal too
weak
The pulse signals on the cuff are too
weak. Re-position the cuff and repeat
the measurement.*
«
ERR 2
» Error signal
During the measurement, error signals
were detected by the cuff, caused for
instance by movement or muscle
tension. Repeat the measurement,
keeping your arm still.
«
ERR 3
» No pressure
in the cuff
An adequate pressure cannot be gener-
ated in the cuff. A leak may have
occurred. Check that the cuff is correctly
connected and is not too loose. Replace
the batteries if necessary. Repeat the
measurement.
«
ERR 5
» Abnormal
result
The measuring signals are inaccurate
and no result can therefore be displayed.
Read through the checklist for
performing reliable measurements and
then repeat the measurement.*
«
HI
»
Pulse or cuff
pressure too
high
The pressure in the cuff is too high (over
300 mmHg) OR the pulse is too high (over
200 beats per minute). Relax for 5
minutes and repeat the measurement.*
«
LO
»
Pulse too
low
The pulse is too low (less than 40 beats
per minute). Repeat the measurement.*
)
If you think the results are unusual, please read through the
information in «Section 1.» carefully.
Summary of Contents for BP A100
Page 35: ...33 BP A100 RU 1 2 3 4 5 6 7 8 9 10 11 12 1 o a e 15 4...
Page 53: ...51 BP A100 BG 1 2 3 4 5 6 7 8 9 10 11 12 1 15 4...
Page 55: ...53 BP A100 BG 4 AM 5 AR 6 3 IV e e ca 7 AO a AO 10 AT AK AL 11 II 12 1 30 1 7 2 4 1 5V AA...
Page 89: ...87 BP A100 GR 1 2 3 4 5 6 7 8 9 10 11 12 1 15 4...
Page 91: ...89 BP A100 GR 4 AM 5 M AR 6 3 IV 7 AO AO 10 AT AK AL 11 II 12 1 ON OFF 30 1 7...
Page 94: ...92 93 42 EEC 3 mmHg 5 4 x 1 5 V AA DC 6V 600 mA EN 1060 1 3 4 IEC 60601 1 IEC 60601 1 2 EMC...
Page 95: ...93 BP A100 GR...
Page 96: ...94 AR Microlife BP A100 2 XXX NJDSPMJGF DPN JDSPMJGF...
Page 97: ...95 BP A100 AR t t t t t t t t t t t t t t t...
Page 98: ...96 z z n z _ _...
Page 99: ...97 BP A100 AR 2 2 2...
Page 100: ...98 n3N2 z n z n z 77 77 77 77 14...
Page 101: ...99 BP A100 AR r r 0 2 0 05 52 r r 3 2...
Page 103: ...101 BP A100 FA r r r r r r r r r r r r r r r r r r r r r r r r r r r r r 3DFHPDNHU r...
Page 104: ...102 PP J 2 PP J 3 7 0 6 s s s s s _ 9 _ 4 r r...
Page 105: ...103 BP A100 FA AR 0 0 0 3 9 AO s 7 r r 1 21 2 AN AL AK AT 21 2 AM...
Page 106: ...104 1L0 9 P 5 55 55 55 55 55 55 0 0 0 0 PP J 2...
Page 107: ...105 BP A100 FA r r r r r r r r r r r r r RVFLOORPHWULF RURWNRII r P 9 r 1 0 LUHFWLYH...