9
* Please consult your doctor, if this or any other problem occurs repeatedly.
Other possible malfunctions and their elimination –
If problems occur when using the device,
the following points should be checked and if necessary, the corresponding measures are to be taken:
☞
Further Information
The level of blood pressure is subject to fluctuations even with healthy people. Important thereby
is, that comparable
measurements always require the same conditions (rest conditions)!
You must consult your specialist dealer or chemist if there are technical problems with the blood
pressure instrument.
Never attempt to repair the instrument yourself!
Any unauthorised opening of the instrument invalidates all guarantee claims!
Error No.
Possible cause(s)
ERR 1
No pulse has been detected.
ERR 2
Unnatural pressure impulses influence the measurement result. Reason: The
arm was moved during the measurement (Artefact).
ERR 3
The inflation of the cuff takes too long. The cuff is not correctly seated.
ERR 5
The measured reading indicated an unacceptable difference between systolic
and diastolic pressures. Take another reading following directions carefully.*
HI
The pressure in the cuff is too high (over 300 mmHg) OR the pulse is too high
(over 200 beats per minute). Relax for 5 minutes and repeat the measurement.*
LO
The pulse is too low (less than 40 beats per minute). Repeat the measurement.*
Malfunction
Remedy
The display remains empty when the
instrument is switched on although the
batteries are in place.
1. Check whether the batteries are installed with
the correct polarity and correct if necessary.
2. If the display is unusual, re-insert batteries or
exchange them.
The device frequently fails to measure the
blood pressure values, or the values
measured are too low (too high).
1. Check the positioning of the cuff.
2. Measure the blood pressure again in complete
peace and quiet under observance of the details
made under point 5.
Every measurement results in different
values although the device functions nor-
mally and the values displayed are normal.
• Please read the following information and the
points listed under «Common sources of error».
Repeat the measurement.
The blood pressure values measured differ
from those measured by the doctor.
• Record the daily development of the values and
consult your doctor about them.
Summary of Contents for BP 3BE0-4
Page 27: ...24 48...
Page 29: ...26 1 1 1 1 2...
Page 30: ...27 2 2 1 2 2 90 140 100 60...
Page 31: ...28 100 60 140 100 90 60 160 140 100 90 180 160 110 100 180 110 120 2 3...
Page 32: ...29 40...
Page 33: ...30 3 19 5 13 5 19 5 13 5 WC1 4 4 1 1 5 2 O I...
Page 34: ...31 1 5 AAA 1 2 4 2 4 3 1 M Time Time M 2 3 Time 4 5 Time...
Page 35: ...32 5 5 1 5 2 15 10 5 3 10...
Page 36: ...33 5 4 O I 75 73 118 5...
Page 38: ...35 HI 300 5 200 LO 40 1 2 1 2 5...
Page 39: ...36 7 8...
Page 41: ...40...
Page 43: ...42 1 1 1 2 1 Pacemaker...
Page 44: ...43 2 1 2 2 2 mmHg 90 mmHg 140 100 mmHg 60 mmHg...
Page 45: ...44 mmHg 100 60 140 100 90 60 160 140 100 90 180 160 110 100 180 110 mmHg 120 3 2...
Page 46: ...45 40...
Page 47: ...46 3 19 5 13 5 19 5 13 5 4 1 4 1 5 LCD Time MEMORY Start O I...
Page 48: ...47 1 5 AAA 1 2 O I 2 4 3 4 1 M Time M 2 3 4 3 5...
Page 49: ...48 5 1 5 5 2 5 15 mmHg 10 3 5...
Page 50: ...49 10 4 5 O I START...
Page 52: ...51 6 2 ERR 1 ERR 2 artefact ERR 3 ERR 5 HI 300 5 200 LO 40 1 2 1 5 2...
Page 53: ...52 7 2...
Page 54: ...53 8 3 9 145 75 70 80 50 5 90 15 40 10 mmHg 280 30 200 40 mmHg 299 0 40 10 15 95 15 95 55 20...