7
•
If the wrist artery lies considerably lower (higher) than the heart, an erroneously higher (lower)
blood-pressure will be measured! (Each 15 cm difference in height results in a measurement
error of 10 mmHg!)
•
A loose cuff causes false measurement values.
5.3. Fitting the cuff
a)
Remove all eventual objects and jewellery (e.g. wrist-
watch) from the wrist in question. Draw the cuff over
the wrist.
b)
The distance between the cuff and the hand should be
approx. 10 mm.
c)
Secure the cuff with the Velcro fastener, so that it lies
comfortably and
not too tight,
whereby no space
should remain between the cuff and the wrist.
d)
Lay the arm on a table, with the palm upwards. Support
the arm a little with a rest (cushion), so that the cuff
rests at about the same height as the heart. Take care,
that the cuff lies free.
5.4. Measuring procedure
After the cuff has been appropriately positioned and the
setting of the pressure pre-selection has been undertaken,
the measurement can begin:
a)
Press O/I button to put the device into operation
b)
Press the START button – the pump begins to inflate the
cuff. In the display, the increasing cuff-pressure is
continually displayed.
c)
After reaching the inflation pressure, the pump stops and the pressure
slowly falls away. The cuff-pressure (large characters) is displayed
during the measurement. When the device has detected the pulse, the
heart symbol in the display begins to blink and a peep tone is audible
for every pulse beat.
Summary of Contents for BP 3BE0-4
Page 27: ...24 48...
Page 29: ...26 1 1 1 1 2...
Page 30: ...27 2 2 1 2 2 90 140 100 60...
Page 31: ...28 100 60 140 100 90 60 160 140 100 90 180 160 110 100 180 110 120 2 3...
Page 32: ...29 40...
Page 33: ...30 3 19 5 13 5 19 5 13 5 WC1 4 4 1 1 5 2 O I...
Page 34: ...31 1 5 AAA 1 2 4 2 4 3 1 M Time Time M 2 3 Time 4 5 Time...
Page 35: ...32 5 5 1 5 2 15 10 5 3 10...
Page 36: ...33 5 4 O I 75 73 118 5...
Page 38: ...35 HI 300 5 200 LO 40 1 2 1 2 5...
Page 39: ...36 7 8...
Page 41: ...40...
Page 43: ...42 1 1 1 2 1 Pacemaker...
Page 44: ...43 2 1 2 2 2 mmHg 90 mmHg 140 100 mmHg 60 mmHg...
Page 45: ...44 mmHg 100 60 140 100 90 60 160 140 100 90 180 160 110 100 180 110 mmHg 120 3 2...
Page 46: ...45 40...
Page 47: ...46 3 19 5 13 5 19 5 13 5 4 1 4 1 5 LCD Time MEMORY Start O I...
Page 48: ...47 1 5 AAA 1 2 O I 2 4 3 4 1 M Time M 2 3 4 3 5...
Page 49: ...48 5 1 5 5 2 5 15 mmHg 10 3 5...
Page 50: ...49 10 4 5 O I START...
Page 52: ...51 6 2 ERR 1 ERR 2 artefact ERR 3 ERR 5 HI 300 5 200 LO 40 1 2 1 5 2...
Page 53: ...52 7 2...
Page 54: ...53 8 3 9 145 75 70 80 50 5 90 15 40 10 mmHg 280 30 200 40 mmHg 299 0 40 10 15 95 15 95 55 20...