45
10. Technical specifications:
Weight
.........................................
145 g (with bat cuff)
Size
...............................................
80 mm x 70 mm x 75 mm (including cuff)
Storage temperature
................. -5 bis 50 ˚C
Humidity
...................................... 15 to 85% relative humidity maximum
Operation temperature
............. 10 to 40 ˚C
Display
.........................................
LCD-Display (Liquid Crystal Display)
Measuring method
.................... oscillometric
Pressure sensor
......................... capacitive
Measuring range SYS/DIA
....... 30 to 280 mmHg
Measuring range Puls
.............. 40 to 200 per minute
Cuff pressure display range
..... 0-299 mmHg
Measuring resolution
............... 1 mmHg
Accuracy
..................................... Pressure within ± 3 mmHg
Pulse ± 5 % of the reading
Memory
........................................ Storing the last 30 measurements automatically
Power source
.............................
2 dry cells (Batteries) UM-4, size AAA 1.5V
Disposal
....................................... Used batteries must be disposed of at a
household hazardous waste collection site.
Defective and out of use devices, must be
returned to their place of purchase.
Accessories
................................ storage case
Technical alterations reserved!
11. www.aponorm.de
Detailed user information about our blood pressure monitors and thermometers
as well as services can be found at www.aponorm.de.
9663_Mobil_GebrAnw_innen_RZ.indd 45
02.12.2008 11:46:03 Uhr
zur Verfügung gestellt von Gebrauchs.info
10. Technical specifications:
10. Technical specifications:
Aktualisierungsstand Gebrauchs.info November 2011