aponorm
®
Basis Control
17
EN
Device care
Clean the device only with a soft, dry cloth.
Cleaning the cuff
Carefully remove spots on the cuff with a damp cloth and
soapsuds.
WARNING:
Do not wash the cuff in a washing machine
or dishwasher!
Accuracy test
We recommend this device is tested for accuracy every 2 years
or after mechanical impact (e.g. being dropped). Please contact
aponorm
®
by microlife-Service to arrange the test (see
foreword).
Disposal
Batteries and electronic devices must be disposed of in
accordance with the locally applicable regulations, not
with domestic waste.
11. Guarantee
This device is covered by a
3 year guarantee
from the date of
purchase. The guarantee is valid only on presentation of the
guarantee card completed by the dealer (see back) confirming
date of purchase or the receipt.
• Batteries and wearing parts are not included.
• Opening or altering the device invalidates the guarantee.
• The guarantee does not cover damage caused by improper
handling, discharged batteries, accidents or non-compliance
with the operating instructions.
• The cuff is included for the functional guarantee only
(tightness) for 2 years.
Please contact
aponorm
®
by microlife-Service (see foreword).
12. Technical Specifications
Operating conditions:
10 - 40 °C / 50 - 104 °F
15 - 95 % relative maximum humidity
Storage conditions:
-20 - +55 °C / -4 - +131 °F
15 - 95 % relative maximum humidity
Weight:
340 g (including batteries)
Dimensions:
135,5 x 82 x 57 mm
Measuring procedure:
oscillometric, corresponding to
Korotkoff method: Phase I systolic,
Phase V diastolic
Measurement range:
20 - 280 mmHg – blood pressure
40 - 200 beats per minute – pulse
Cuff pressure display
0 - 299 mmHg
range:
Resolution:
1 mmHg
Static accuracy:
pressure within ± 3 mmHg
Pulse accuracy:
± 5 % of the readout value
Voltage source:
• 4 x 1.5 V Alkaline-Batteries; size
AA
• Mains adapter DC 6 V, 600 mA
(optional)
Battery life:
approx. 920 measurements
(new
batteries)
IP class:
IP
20
Reference to
EN 1060-1 /-3 /-4; IEC 60601-1-2
standards:
(EMC); IEC 60601-1-11
Expected service life:
Device: 5 years or 10,000
measurements; Accessories: 2 years
This device complies with the requirements of the Medical
Device Directive 93/42/EEC.
Technical alterations reserved.
Summary of Contents for Aponorm 06575428
Page 1: ...Gebrauchsanweisung DE 2 RU 26 EN 10 TR 18 Oberarm B A S I S C O N T R O L...
Page 3: ...aponorm by 1 DE 1 2 3 4 5 6 7 8 10 11 12 13 14 15 16 17 18 19 20 9 aponorm Basis Control...
Page 29: ...aponorm Basis Control 27 RU...
Page 30: ...aponorm by 28 6 6 140 80 130 90 5 9 8 0 z 100 160 y z 60 100 y...
Page 31: ...aponorm Basis Control 29 RU aponorm PLFUROLIH 1 16 12 13 14 6 0 0 aponorm PLFUROLIH 7 3...
Page 32: ...aponorm by 30 1 M 20 19 15 6 6 9 M 20 0...
Page 33: ...0 10 10 5 aponorm Basis Control 31 RU 1L0...
Page 34: ...aponorm by 32 aponorm PLFUROLIH aponorm PLFUROLIH 4 55 55 17 55 2 55 55 18...
Page 35: ...DSRQRUP PLFUROLIH aponorm Basis Control 33 RU...
Page 36: ...aponorm by 34 aponorm PLFUROLIH 9 9...