Page 11
Overview
The SET menu is similar in operation to the MAIN MENU, however the SET menu is used to set
the calibration for the unit and the unit of measure. See the section about operation (setting the
scale) for additional information on using the SET menu. Below, each menu is described in
detail. Calibration only needs to be set for one axis to calibrate all directions.
Set Scale Using X Axis
This will set the calibration using the horizontal (left to right) axis, or the distance between two
vertical (up-down) lines. It will accept any number, up to four digits between 0.0001 and 9999 for
the calibration.
Note that measuring is much more precise when setting the calibration on the X axis, as
opposed to the Y axis. Calibration is also more precise when the distance between the two lines
is as wide as possible during calibration.
Set Scale Using Y Axis
This will set the calibration Using the vertical (top to bottom) axis, or the distance between two
horizontal (left-right) lines. Enter any number, up to four digits between 0.0001 and 9999 for the
calibration.
Note that measuring is much more precise when setting the calibration on the X axis. Y axis
calibration is provided for those places where X axis calibration is impractical. Calibration is also
more precise when the distance between the two lines is as wide as possible during calibration.
CALIBRATION SET
Scale and Unit Parameters
1 Set Scale Using X Axis
2 Set Scale Using Y Axis
3 Clear Scale
4 Select Unit
5 Define Custom Unit
0 Exit
Press number for selection.