2.2.5.12 - LCD – Display and Buzzer operation
Stato
(Status)
→
Non Disp.
Non Disp. (
No Parameters
)
Opzioni
(Options)
→
Bkl
Auto
[Auto / On]
Livelli
(Trip Level)
→
Non Disp.
Non Disp. (
No Parameters
)
Tempi
(Timer)
→
Non Disp.
Non Disp. (
No Parameters
)
Backlight.
ON
= Backlight ON.
Bkl
:
Auto
= Switched-on Automatically on operation of Keyboard buttons. Off
after 10sec of inactivity.
2.2.5.13 - Comm – Communication Parameters
Stato
(Status)
→
Non Disp.
Non Disp. (
No Parameters
)
Opzioni
(Options)
→
LBd
9600
[9600 / 19200 / 38400 / 57600]
→
RBd
9600
[9600 / 19200]
→
Mod
8,n,1
[8,n,1 / 8,o,1 / 8,e,1]
→
RPr
Modbus
[Modbus]
Livelli
(Trip Level)
→
Non Disp.
Non Disp. (
No Parameters
)
Tempi
(Timer)
→
Non Disp.
Non Disp. (
No Parameters
)
LBd
:
Local Baud Rate (Front panel RS232 communication speed)
RBd
:
Remote Baud Rate (Rear panel terminal blocks RS485 communication speed)
Mod
:
Communication mode (communication parameters)
Note: Any change of this setting becomes valid at the next power on
RPr
:
Remote Protocol