MicroAire ASP-1021 Instructions For Use Manual Download Page 9

IM-LIPOTOWER Rev G

9

CONNECTING A SALINE/FLUID BAG TO THE 
TUMESCENT POLE

1.  Hang a sterile saline/fluid bag onto the 

Tumescnt Pole by carefully inserting the 
bag’s hanging-grommet onto one of the 
four hooks on the Tumescent Pole 

(Figure 

17 and 18).  

Up to four bags can be hung 

at one time, with one on each hook.

WARNING: 

Ensure that saline bags are properly secured to the hook.

WARNING: 

Tumescent pole maximum safe weight capacity is 3kg (6.6lb) per hook. Do not exceed.

CONNECTING INFILTRATION TUBING

1.  Attach sterile, single-spike infiltration 

tubing (REF ASP-TB-TUM) to the saline/
fluid bag according to the instructions 
for the saline/fluid bag.

2.  Lift the lid on the Tumescent Pump 

located on the side of the cart. 

(Figure 19)

3.  Locate the soft silicone section of 

infiltration tubing and place it into the 
open pump, ensuring that the outward 
flow of fluid is exiting the pump towards the front of the cart.  

(Figure 20)

4.  Close the lid on the Tumescent Pump to compress the Infiltration Tubing within the pump. 

OPERATION

WARNING:

 Prior to use, all MicroAire Power Assisted Lipoplasty System subsystem manuals should be reviewed for 

important warnings and instructions for use.

WARNING: 

Prior to use, all system components should be inspected to detect any damage, excessive wear, corrosion 

or malfunction. DO NOT use if any potential problems become apparent.  Contact 
MicroAire for resolution.

WARNING:

 Eye protection must be worn when operating any power equipment. 

Dislodged burs, blades or bone fragments can result in eye injury, blindness, or 
contamination of the eye from patient tissue or body fluids.

POWERING UP THE SYSTEM

1.  Before plugging the wall outlet power cord into a hospital-grade wall 

outlet, ensure that the STANDBY switch on the front of the cart is in the 
“STANDBY”  ( 

o

 ) position. 

(Figure 21)

2.  Insert the electrical cord into the POWER INLET, located on the lower left 

of the rear of the cart 

(Figure 22)

. Plug the power cord into an appropriate 

hospital-grade wall outlet. 

3.  To power up the MicroAire LipoTower System, push the Mains Power 

Disconnect Switch to the “ON”  ( 

|

 ) position 

(Figure 22). 

 A visible light on 

the STANDBY switch will now be illuminated.

  

4.  Push the STANDBY switch to the “ON”  ( 

|

 ) position. An audible sound 

from the cooling fans will indicate that the MicroAire LipoTower System is 
powered up.

Figure 18 - Make surethe bag is se-
curely attached to the hook.

Figure 21 - Standby switch

Figure 22 - Power inlet and Mains 
Disconnect switch

Figure 20 - Insert Infiltration Tubing

Figure 19 - Open the Tumescent Pump

Figure 17 - Attach saline/fluid bag to 
the Tumescent Pole

Summary of Contents for ASP-1021

Page 1: ...MicroAire LipoTower System Instructions for Use...

Page 2: ...nsole for PAL REF 5025 Electric Power Console for MicroAire Instruments REF PAL 1200 Liposuction Tubing 12 foot REF 1006 PALE PAL Instrument Cable for REF 1020 1025 REF 5006 PAL PAL Instrument Cable f...

Page 3: ...nd the 5020 5025 Electric Console are described in their respective IFUs This IFU addresses primarily the operation maintenance cleaning and sterilization procedures applicable to the MicroAire LipoTo...

Page 4: ...nd or postoperative hemodynamic stability and patient safety The capability of providing adequate timely replacement is essential for patient safety WARNING Grounding reliability can only be achieved...

Page 5: ...system on another patient CAUTION Federal Law restricts this device to sale by or on the order of a physician or properly licensed practitioner CAUTION See PAL 650 Instrument System Instruction Manual...

Page 6: ...CH ON allows pumps to be activated by the Control Switch STANDBY o turns vacuum pumps off 6 STORAGE BASKET 7 LOCKING CASTERS 8 PUSH HANDLE 9 MAIN VACUUM CONNECTION 10 REAR STORAGE BASKET 11 POWER INLE...

Page 7: ...20 1025 or REF 5020 through the cable opening and connect it to the hospital grade outlet on the back of the cart Figure 6 4 Manuever the BNC cable from the REF 1020 1025 or REF 5020 through the cable...

Page 8: ...er Rack on the top of the cart Slide one or more Waste Canister Rings REF ASP RNG 1200 or REF ASP RNG 2000 into any of the four available canister slots These slots also accept various wall mount styl...

Page 9: ...or to use all system components should be inspected to detect any damage excessive wear corrosion or malfunction DO NOT use if any potential problems become apparent Contact MicroAire for resolution W...

Page 10: ...PRESSURE ADJUSTMENT knob controls the amount of suction pressure Vacuum pressure is displayed on the LCD screen Suction is increased by turning the knob clockwise towards MAX and suction is decreased...

Page 11: ...N OPTIONAL INSTALLING THE TUMESCENT MODULE 1 Turn off power and disconnect main power by unplugging power cord 2 Locate the Tumescent Bay cover on the left side of the cart Remove the the two retainin...

Page 12: ...the foot switch still does not operate properly contact MicroAire Customer Service for additional troubleshooting and repair VACUUM PRESSURE IS INSUFFICIENT Ensure all aspiration canisters and aspira...

Page 13: ...roAire LipoTower System is operated at high elevation or low atmospheric pressure aspiration performance may be adversely affected NOISE LEVEL EXPOSURE LIMITS The A weighted emission sound pressure le...

Page 14: ...SUMMARY CAUTION The MicroAire LipoTower system needs to be installed and put into service according to the EMC information provided in this Instructions for Use manual Portable and mobile RF communica...

Page 15: ...IM LIPOTOWER Rev G 15 Table 1 IEC 60601 1 2 Table 201 Table 2 IEC 60601 1 2 Table 202...

Page 16: ...IM LIPOTOWER Rev G 16 Table 4 IEC 60601 1 2 Table 206 Table 3 IEC 60601 1 2 Table 204...

Page 17: ...IM LIPOTOWER Rev G 17 NOTES...

Page 18: ...IM LIPOTOWER Rev G 18 NOTES...

Page 19: ...IM LIPOTOWER Rev G 19 NOTES...

Page 20: ...nstruments LLC IM LIPOTOWER Rev G Printed in the USA M MicroAire Surgical Instruments LLC 3590 Grand Forks Blvd Charlottesville VA 22911 USA Telephone 800 722 0822 434 975 8000 Fax 800 648 4309 434 97...

Reviews: