MicroAire ASP-1021 Instructions For Use Manual Download Page 11

IM-LIPOTOWER Rev G

11

MONITORING

Primary monitoring of the MicroAire LipoTower System status will be accomplished by observing progress at the 
procedure site of the patient. Secondary monitoring of the MicroAire LipoTower System status will be accomplished 
by periodic viewing of the tumescent fluid bags, the Tumescent Measuring Device (where applicable), the vacuum 
pressure as displayed on the LCD screen, and by viewing the waste canisters, to verify functionality as well as volume 
of aspirant.

CAUTION: 

Waste canisters should be replaced when 90% full. Overfilling may result in aspirant matter getting into, 

and contaminating, the MicroAire LipoTower System.

Once the MicroAire LipoTower System is powered up, and its appropriate cables, tubing, waste canisters and biofilter 
properly connected, and its mode of operation selected, and once the REF PAL-650 or PAL-600 and the REF 1020/1025 
or REF 5020 have been properly powered and connected in accordance with their respective IFUs, surgical procedures 
can be started.

CAUTION: 

Before use, verify that the MicroAire LipoTower System functions properly. If not, contact MicroAire for 

resolution.

ADDITIONAL SET UP AND INSTALLATION (OPTIONAL)

INSTALLING THE TUMESCENT MODULE

1.  Turn off power and disconnect main 

power by unplugging power cord.

2.  Locate the Tumescent Bay cover on 

the left side of the cart. Remove the 
the two retaining screws and slide the 
cover towards the back of the cart to 
disengage the clip that is near the front 
of the cart. Remove cover. 

(Figure 29)

3.  Locate the electrical connector harness 

on the pump module and connect it to 
the corresponding electrical connector 
within the bay.

 (Figure 30)

4.  Insert the Tumescent Module (REF 

ASP-TUM) into the tumescent bay by 
inserting the pump module and then 
sliding it towards the front of the cart to 
engage the retaining clip. 

(Figure 31)

5.  Insert the two retaining screws to secure 

the module in the bay. Do not tighten 
the screws until both are in place, to 
allow for adjusting the levelness of the 
module. 

(Figure 32)

6.  Align the slot on the peristaltic pump 

head with the drive shaft, then attach 
the pump head. 

(Figure 33)

7.  Rotate the pump head clockwise until it 

clicks into place. 

(Figure 34)

Figure 29 - Remove screws and slide 
cover towards the back of the cart.

Figure 30 - Connect the electrical 
harness.

Figure 32 - Replace retaining screws.

Figure 31 - Insert the module. Slide it 
rightward to engage retaining clip.

Figure 34 - Rotate the pump head 
clockwise until it clicks into place.

Figure 33 - Align the slot on the pump 
head with the drive shaft.

Summary of Contents for ASP-1021

Page 1: ...MicroAire LipoTower System Instructions for Use...

Page 2: ...nsole for PAL REF 5025 Electric Power Console for MicroAire Instruments REF PAL 1200 Liposuction Tubing 12 foot REF 1006 PALE PAL Instrument Cable for REF 1020 1025 REF 5006 PAL PAL Instrument Cable f...

Page 3: ...nd the 5020 5025 Electric Console are described in their respective IFUs This IFU addresses primarily the operation maintenance cleaning and sterilization procedures applicable to the MicroAire LipoTo...

Page 4: ...nd or postoperative hemodynamic stability and patient safety The capability of providing adequate timely replacement is essential for patient safety WARNING Grounding reliability can only be achieved...

Page 5: ...system on another patient CAUTION Federal Law restricts this device to sale by or on the order of a physician or properly licensed practitioner CAUTION See PAL 650 Instrument System Instruction Manual...

Page 6: ...CH ON allows pumps to be activated by the Control Switch STANDBY o turns vacuum pumps off 6 STORAGE BASKET 7 LOCKING CASTERS 8 PUSH HANDLE 9 MAIN VACUUM CONNECTION 10 REAR STORAGE BASKET 11 POWER INLE...

Page 7: ...20 1025 or REF 5020 through the cable opening and connect it to the hospital grade outlet on the back of the cart Figure 6 4 Manuever the BNC cable from the REF 1020 1025 or REF 5020 through the cable...

Page 8: ...er Rack on the top of the cart Slide one or more Waste Canister Rings REF ASP RNG 1200 or REF ASP RNG 2000 into any of the four available canister slots These slots also accept various wall mount styl...

Page 9: ...or to use all system components should be inspected to detect any damage excessive wear corrosion or malfunction DO NOT use if any potential problems become apparent Contact MicroAire for resolution W...

Page 10: ...PRESSURE ADJUSTMENT knob controls the amount of suction pressure Vacuum pressure is displayed on the LCD screen Suction is increased by turning the knob clockwise towards MAX and suction is decreased...

Page 11: ...N OPTIONAL INSTALLING THE TUMESCENT MODULE 1 Turn off power and disconnect main power by unplugging power cord 2 Locate the Tumescent Bay cover on the left side of the cart Remove the the two retainin...

Page 12: ...the foot switch still does not operate properly contact MicroAire Customer Service for additional troubleshooting and repair VACUUM PRESSURE IS INSUFFICIENT Ensure all aspiration canisters and aspira...

Page 13: ...roAire LipoTower System is operated at high elevation or low atmospheric pressure aspiration performance may be adversely affected NOISE LEVEL EXPOSURE LIMITS The A weighted emission sound pressure le...

Page 14: ...SUMMARY CAUTION The MicroAire LipoTower system needs to be installed and put into service according to the EMC information provided in this Instructions for Use manual Portable and mobile RF communica...

Page 15: ...IM LIPOTOWER Rev G 15 Table 1 IEC 60601 1 2 Table 201 Table 2 IEC 60601 1 2 Table 202...