MONITORING
Primary monitoring of the MicroAire LipoTower System status will be accomplished by observing progress at the
procedure site of the patient. Secondary monitoring of the MicroAire LipoTower System status will be accomplished
by periodic viewing of the tumescent fluid bags, the Tumescent Measuring Device (where applicable), the vacuum
pressure as displayed on the LCD screen, and by viewing the waste canisters, to verify functionality as well as volume
of aspirant.
CAUTION:
Waste canisters should be replaced when 90% full. Overfilling may result in aspirant matter getting into,
and contaminating, the MicroAire LipoTower System.
Once the MicroAire LipoTower System is powered up, and its appropriate cables, tubing, waste canisters and biofilter
properly connected, and its mode of operation selected, and once the REF PAL-650 or PAL-600 and the REF 1020/1025
or REF 5020 have been properly powered and connected in accordance with their respective IFUs, surgical procedures
can be started.
CAUTION:
Before use, verify that the MicroAire LipoTower System functions properly. If not, contact MicroAire for
resolution.
ADDITIONAL SET UP AND INSTALLATION (OPTIONAL)
INSTALLING THE TUMESCENT MODULE
1. Turn off power and disconnect main
power by unplugging power cord.
2. Locate the Tumescent Bay cover on
the left side of the cart. Remove the
the two retaining screws and slide the
cover towards the back of the cart to
disengage the clip that is near the front
of the cart. Remove cover.
(Figure 29)
3. Locate the electrical connector harness
on the pump module and connect it to
the corresponding electrical connector
within the bay.
(Figure 30)
4. Insert the Tumescent Module (REF
ASP-TUM) into the tumescent bay by
inserting the pump module and then
sliding it towards the front of the cart to
engage the retaining clip.
(Figure 31)
5. Insert the two retaining screws to secure
the module in the bay. Do not tighten
the screws until both are in place, to
allow for adjusting the levelness of the
module.
(Figure 32)
6. Align the slot on the peristaltic pump
head with the drive shaft, then attach
the pump head.
(Figure 33)
7. Rotate the pump head clockwise until it
clicks into place.
(Figure 34)
Figure 29 - Remove screws and slide
cover towards the back of the cart.
Figure 30 - Connect the electrical
harness.
Figure 32 - Replace retaining screws.
Figure 31 - Insert the module. Slide it
rightward to engage retaining clip.
Figure 34 - Rotate the pump head
clockwise until it clicks into place.
Figure 33 - Align the slot on the pump
head with the drive shaft.
Summary of Contents for ASP-1021
Page 1: ...MicroAire LipoTower System Instructions for Use...
Page 15: ...IM LIPOTOWER Rev G 15 Table 1 IEC 60601 1 2 Table 201 Table 2 IEC 60601 1 2 Table 202...
Page 16: ...IM LIPOTOWER Rev G 16 Table 4 IEC 60601 1 2 Table 206 Table 3 IEC 60601 1 2 Table 204...
Page 17: ...IM LIPOTOWER Rev G 17 NOTES...
Page 18: ...IM LIPOTOWER Rev G 18 NOTES...
Page 19: ...IM LIPOTOWER Rev G 19 NOTES...