MicroAire ASP-1021 Instructions For Use Manual Download Page 10

IM-LIPOTOWER Rev G

10

TUMESCENT INFILTRATION CONTROLS

1.  Turn on the Tumescent Pump using the 

ON/OFF switch on the side of the cart. 

(Figure 23)

2.  Adjust the speed/flow rate of the pump 

by turning the Speed Adjustment Knob. 
Flow rate is increased by turning the 
knob clockwise, and speed is decreased 
by turning the knob counterclockwise 

(Figure 24)

ASPIRATION CONTROLS

1.  The 3-WAY CONTROL SWITCH controls 

the ON/OFF status of the MicroAire 
LipoTower System vacuum pumps. 

(Figure 25)

• 

BYPASS (left position) makes the 
vacuum pumps continuously active; 

• 

PAL (middle position) places the ON/
OFF status of the pumps under the 
control of the throttle of the MicroAire PAL handpiece; 

• 

FOOT SWITCH (right position) places the ON/OFF status of the pumps under the control of a foot switch.

NOTE: 

If “FOOT SWITCH” is selected, ensure that the foot switch cable is securely connected to its receptacle located 

on the lower rear of the MicroAire LipoTower System. Ensure that the foot switch pedal is located within reach of the 
operator. 

2.  The VACUUM PRESSURE ADJUSTMENT knob controls the amount of suction pressure.  Vacuum pressure is 

displayed on the LCD screen. Suction is increased by turning the knob clockwise towards MAX, and suction is 
decreased by turning the knob counterclockwise towards MIN. 

(Figure 26)

LCD TOUCH SCREENS

BOOT-UP SCREEN

The MicroAire logo will appear while the system 
is starting up.  After boot-up, the screen will 
automatically advance to the home screen.

HOME SCREEN

This screen displays vacuum pressure, and an option to 
select the Tools screen. 

(Figure 27)

TOOLS SCREEN

This screen shows the number of hours used by the Biofilter, a button to reset the Biofilter to Zero, the number of total 
hours used by the vacuum pump, and buttons to select the unit of measurement. 

(Figure 28)

• 

Reset the biofilter hour meter by selecting RESET BIOFILTER TO ZERO.

• 

Select preferred units of measurement by choosing “INCH” or “MM.”

• 

To return to the Main Screen, select the Back icon (left-facing arrow).

NOTE: 

 The vacuum hour meter is not resettable, it represents total time used by the vacuum pumps.

Figure 24 - Speed Adjustment knob

Figure 23 - Tumescent Pump ON/OFF 

Figure 25 - Three-Way Control Switch

Figure 26 - Vacuum Pressure Knob

Figure 27 - Home Screen

Figure 28 - Tools Screen

Summary of Contents for ASP-1021

Page 1: ...MicroAire LipoTower System Instructions for Use...

Page 2: ...nsole for PAL REF 5025 Electric Power Console for MicroAire Instruments REF PAL 1200 Liposuction Tubing 12 foot REF 1006 PALE PAL Instrument Cable for REF 1020 1025 REF 5006 PAL PAL Instrument Cable f...

Page 3: ...nd the 5020 5025 Electric Console are described in their respective IFUs This IFU addresses primarily the operation maintenance cleaning and sterilization procedures applicable to the MicroAire LipoTo...

Page 4: ...nd or postoperative hemodynamic stability and patient safety The capability of providing adequate timely replacement is essential for patient safety WARNING Grounding reliability can only be achieved...

Page 5: ...system on another patient CAUTION Federal Law restricts this device to sale by or on the order of a physician or properly licensed practitioner CAUTION See PAL 650 Instrument System Instruction Manual...

Page 6: ...CH ON allows pumps to be activated by the Control Switch STANDBY o turns vacuum pumps off 6 STORAGE BASKET 7 LOCKING CASTERS 8 PUSH HANDLE 9 MAIN VACUUM CONNECTION 10 REAR STORAGE BASKET 11 POWER INLE...

Page 7: ...20 1025 or REF 5020 through the cable opening and connect it to the hospital grade outlet on the back of the cart Figure 6 4 Manuever the BNC cable from the REF 1020 1025 or REF 5020 through the cable...

Page 8: ...er Rack on the top of the cart Slide one or more Waste Canister Rings REF ASP RNG 1200 or REF ASP RNG 2000 into any of the four available canister slots These slots also accept various wall mount styl...

Page 9: ...or to use all system components should be inspected to detect any damage excessive wear corrosion or malfunction DO NOT use if any potential problems become apparent Contact MicroAire for resolution W...

Page 10: ...PRESSURE ADJUSTMENT knob controls the amount of suction pressure Vacuum pressure is displayed on the LCD screen Suction is increased by turning the knob clockwise towards MAX and suction is decreased...

Page 11: ...N OPTIONAL INSTALLING THE TUMESCENT MODULE 1 Turn off power and disconnect main power by unplugging power cord 2 Locate the Tumescent Bay cover on the left side of the cart Remove the the two retainin...

Page 12: ...the foot switch still does not operate properly contact MicroAire Customer Service for additional troubleshooting and repair VACUUM PRESSURE IS INSUFFICIENT Ensure all aspiration canisters and aspira...

Page 13: ...roAire LipoTower System is operated at high elevation or low atmospheric pressure aspiration performance may be adversely affected NOISE LEVEL EXPOSURE LIMITS The A weighted emission sound pressure le...

Page 14: ...SUMMARY CAUTION The MicroAire LipoTower system needs to be installed and put into service according to the EMC information provided in this Instructions for Use manual Portable and mobile RF communica...

Page 15: ...IM LIPOTOWER Rev G 15 Table 1 IEC 60601 1 2 Table 201 Table 2 IEC 60601 1 2 Table 202...

Page 16: ...IM LIPOTOWER Rev G 16 Table 4 IEC 60601 1 2 Table 206 Table 3 IEC 60601 1 2 Table 204...

Page 17: ...IM LIPOTOWER Rev G 17 NOTES...

Page 18: ...IM LIPOTOWER Rev G 18 NOTES...

Page 19: ...IM LIPOTOWER Rev G 19 NOTES...

Page 20: ...nstruments LLC IM LIPOTOWER Rev G Printed in the USA M MicroAire Surgical Instruments LLC 3590 Grand Forks Blvd Charlottesville VA 22911 USA Telephone 800 722 0822 434 975 8000 Fax 800 648 4309 434 97...

Reviews: