IM-LIPOTOWER Rev G
10
TUMESCENT INFILTRATION CONTROLS
1. Turn on the Tumescent Pump using the
ON/OFF switch on the side of the cart.
(Figure 23)
2. Adjust the speed/flow rate of the pump
by turning the Speed Adjustment Knob.
Flow rate is increased by turning the
knob clockwise, and speed is decreased
by turning the knob counterclockwise
(Figure 24)
ASPIRATION CONTROLS
1. The 3-WAY CONTROL SWITCH controls
the ON/OFF status of the MicroAire
LipoTower System vacuum pumps.
(Figure 25)
•
BYPASS (left position) makes the
vacuum pumps continuously active;
•
PAL (middle position) places the ON/
OFF status of the pumps under the
control of the throttle of the MicroAire PAL handpiece;
•
FOOT SWITCH (right position) places the ON/OFF status of the pumps under the control of a foot switch.
NOTE:
If “FOOT SWITCH” is selected, ensure that the foot switch cable is securely connected to its receptacle located
on the lower rear of the MicroAire LipoTower System. Ensure that the foot switch pedal is located within reach of the
operator.
2. The VACUUM PRESSURE ADJUSTMENT knob controls the amount of suction pressure. Vacuum pressure is
displayed on the LCD screen. Suction is increased by turning the knob clockwise towards MAX, and suction is
decreased by turning the knob counterclockwise towards MIN.
(Figure 26)
LCD TOUCH SCREENS
BOOT-UP SCREEN
The MicroAire logo will appear while the system
is starting up. After boot-up, the screen will
automatically advance to the home screen.
HOME SCREEN
This screen displays vacuum pressure, and an option to
select the Tools screen.
(Figure 27)
TOOLS SCREEN
This screen shows the number of hours used by the Biofilter, a button to reset the Biofilter to Zero, the number of total
hours used by the vacuum pump, and buttons to select the unit of measurement.
(Figure 28)
•
Reset the biofilter hour meter by selecting RESET BIOFILTER TO ZERO.
•
Select preferred units of measurement by choosing “INCH” or “MM.”
•
To return to the Main Screen, select the Back icon (left-facing arrow).
NOTE:
The vacuum hour meter is not resettable, it represents total time used by the vacuum pumps.
Figure 24 - Speed Adjustment knob
Figure 23 - Tumescent Pump ON/OFF
Figure 25 - Three-Way Control Switch
Figure 26 - Vacuum Pressure Knob
Figure 27 - Home Screen
Figure 28 - Tools Screen
Summary of Contents for ASP-1021
Page 1: ...MicroAire LipoTower System Instructions for Use...
Page 15: ...IM LIPOTOWER Rev G 15 Table 1 IEC 60601 1 2 Table 201 Table 2 IEC 60601 1 2 Table 202...
Page 16: ...IM LIPOTOWER Rev G 16 Table 4 IEC 60601 1 2 Table 206 Table 3 IEC 60601 1 2 Table 204...
Page 17: ...IM LIPOTOWER Rev G 17 NOTES...
Page 18: ...IM LIPOTOWER Rev G 18 NOTES...
Page 19: ...IM LIPOTOWER Rev G 19 NOTES...