MICRO-EPSILON surfaceCONTROL 3D SC2500 Assembly Instructions Manual Download Page 9

Page 9

surfaceCONTROL 3D SC2500 / SC2510

10. 

Operating Programs

Various programs are provided for operation of the sensor. 

 - 3DInspect solves industrial 3D measurement tasks.
 - You can use surfaceCONTROL DefMap3D, including the dongle which is subject to a 

charge, for the analysis of individual surfaces with surfaceCONTROL 3D 25x0.

11. 

Installation

Requirements

The following minimum system specifications are necessary for the use of the operating 
programs:

 - Windows 10 (64 bit)
 -

≥ 2

 GHz processor (64 Bit) or higher

 - 4 GB RAM
 - Screen resolution: 1280 x 1024
 - Graphics card/GPU with OpenGL 3.1 or higher

The following procedure is necessary in order to install the software:
1.  Install the hardware of the Ethernet interface(s) if not present.
2.  Install the software.
3.  With DefMap3D: use the USB dongle required for this.
4.  Connect the controller of the surfaceCONTROL 3D 25x0 measuring system to the PC 
 

via Ethernet. Do not use hubs.

i

  If your network card supports the VLAN option, this option must be disabled.

Connecting surfaceCONTROL 3D 25x0 to the PC

To connect surfaceCONTROL 3D 25x0 via Ethernet to the PC, proceed as follows:
1.  Complete the installation of the software.
2.  Connect surfaceCONTROL 3D 25x0 to the PC using the Ethernet interface.
3.  Turn on the power supply.
3.  Wait until the surfaceCONTROL 3D 25x0 measuring system is recognized by the PC.   
 

This takes about 1 minute.

i

  To be able to operate the measuring system, the controller and the PC must be 

located in the same subnet.

The sensor starts in DHCP mode. If there is no DHCP server in the network or if no ad-
dress has been assigned manually, the sensor and network adapter assign themselves an 
address (Link Local Address).
You can now operate the surfaceCONTROL 3D 25x0 measuring system using the operat-
ing programs listed above.

Summary of Contents for surfaceCONTROL 3D SC2500

Page 1: ...the controller Do not touch the protective windows Remove any fingerprints immediately using pure alcohol and a clean cotton cloth without leaving any streaks Protect the cables against damage Failur...

Page 2: ...Controller 40 85 C Humidity 20 80 non condensing Ambient pressure Atmospheric pressure 1 Max permissible operating temperature with 0 7 m s air blow 4 Unpacking Included in Delivery 1 Sensor surfaceCO...

Page 3: ...uded in risk group 2 according to EN 62471 2008 Do not look into the lens Consciously close your eyes or im mediately turn away if the optical radiation enters the eye The warning sign below is attach...

Page 4: ...e connector panel on the rear side 1 2 3 4 1 Ethernet port Camera 1 2 Ethernet port Camera 2 3 Sensor control 4 Power supply Fig 2 Rear view of sensor with connectors Supply Voltage Power Pin Signal C...

Page 5: ...Pin assignment connector Ethernet port For the image data output of the cameras via Gigabit Ethernet the sensor has two Gigabit Ethernet interfaces The sensor supports only the transmission with 1 Gbi...

Page 6: ...roller front view with connectors Supply voltage Power The power supply connection is a 4 pin plug connector Pin Designation Comment 9 V 36 V DC nominal value 24 V max power consumption 112 W 0 V Fig...

Page 7: ...Dismantling Proceed as follows if you would like to detach the sensor from the tripod again Only for column stand option with boom Secure the boom of the tripod with the help of the clamping lever or...

Page 8: ...the delivery Connect the controller to the PC To do so use a common Ethernet cable of category CAT5e or higher Connect the sensor to the power supply Connect the controller to the power supply Turning...

Page 9: ...red for this 4 Connect the controller of the surfaceCONTROL 3D 25x0 measuring system to the PC via Ethernet Do not use hubs i If your network card supports the VLAN option this option must be disabled...

Page 10: ...the surface of the test object Alignment with Crosshair You will achieve the best results in the mid of the measurement area focal range of cam eras In the main view switch from the 3D View to the Im...

Page 11: ...ere created with the 3DInspect software from Micro Epsilon 14 Disclaimer All components of the device have been checked and tested for functionality in the factory However should any defects occur des...

Page 12: ...ess is defec tive If possible save the current sensor settings in a parameter set to reload them into the sensor after the repair Please send us the affected parts for repair or exchange If the cause...

Reviews: