Page 17
Installation
surfaceCONTROL 2500
5.2.3
Installation Instructions
surfaceCONTROL 2500
surfaceCONTROL 2500/ST
For all connection cables, use only the
appropriate cables from the accessories
As connection cable, use fiber optics with
an OM class of at least OM2 or the corres-
ponding cables from the accessories
.
Lay all connection cables in accordance with the generally applicable measuring tech-
nology regulations, i.e. for example not directly next to pulse-carrying lines, preferably
in a separate cable duct.
The minimum bending radii of the recommended cables for flexible laying must not be
less than 80 mm.
Only use the included power pack for the power supply of the sensor.
> Damage to or destruction of the sensor