Error Codes
scanCONTROL 29xx
12.
Error Codes
(— LED
state
lights for a long time lang, • LED
state
lights briefly)
Flashing
sequence
Cause
Remedy
Notes
Group: Loading / saving configuration
..
2x short
Mode not found.
Select different one.
Only previously stored
modes can be called up.
..-
2x short, 1x long
White error flash
Contact manufacturer, return
device.
Should not occur in normal
operation.
…
3x short
Flash full
None, contact manufacturer.
Should not occur in normal
operation.
….
4x short
Loading suppressed due to
active data transmission.
Stop active data transmission.
Prevents PC software
crashes.
Group: Data processing and transmission
--
2x long
Data overflow in the sensor
Select smaller measuring field,
reduce profile frequency, se-
lect less complex measuring
program.
Data can be impaired; ex-
posure time can be longer
than expected.
--.
2x long, 1x short
Data overflow during receipt
of the data from the sensor.
Select smaller measuring field,
reduce profile frequency, se-
lect less complex measuring
program.
Data can be faulty.
--..
2x long, 2x short
Data overflow for serial port
RS422
Reduce profile frequency, se-
lect less complex measuring
program.
Data can be faulty.