Page 23
Mounting and Installation
scanCONTROL 29xx
5.
Mounting and Installation
5.1
Attachment and Mounting
- using 2 or 3 screws M5, screwed directly
- using 2 or 3 screws M4, screwed pushed through
Depending on the installation position it is recommended to determine the position of the sensor for example
by adjusting screws on the specially marked attachment points.
The pin hole ø3H11 is intended for a position locking pin. The sensor can be mounted reproducible and
replaceable together with an attachment point
, see Fig. 12, see Fig. 13, see Fig. 14.
The mounting dimensions refer to the dimensional drawings.
Pay attention to careful handling during mounting and operation.
> Damage to or destruction of the sensor
The laser beam should strike the target surface at right angles. Otherwise, inaccurate measurements cannot
be excluded.
Mount the sensor by means of screws type M5 or by means of through bores for M4 with the screws
from the accessories.
Bolt connection
Bolt length
Screw
Washer
Torque
33 mm
M4 x ISO 4762-A2 A 4.3 ISO 7089-A2 2 Nm (µ = 0,12)
Direct fastening
Screwing depth
Screw
Torque
min 8 mm, max 10 mm M5 x ISO 4762-A2 3.5 Nm (µ = 0.12)
Fig. 10 Mounting conditions
The bearing surfaces surrounding the fastening holes (through-holes) are slightly raised.
Mount the sensor only to the existing through-holes on a flat surface. Clamps of any kind are not permitted.
Do not exceed torques.
> Inaccurate, erroneous measuring values