Page 30
Mounting and Installation
scanCONTROL 29xx
5.2.2
Power Supply
Connector
pin assignment
Range: 11 V – 30 V (nominal value 24 V) DC; load maximal 500 mA
The supply voltage is protected against reverse polarity.
The cable shield is connected with the connector housing and should be connected to the protective con-
ductor PE of the power supply.
The shielded multifunction cable PCR3000-x is recommended.
The supply voltage for the scanCONTROL 29xx measuring device should come from a 24 V power supply,
which is only used for measuring equipment and not simultaneously for drives, contactors or similar pulse
interference sources. Use a power supply with galvanic isolation.
The multifunction
socket must only
be connected when
the power supply is
switched off.
> Damage to or
destruction of the
sensor