Selection of Measurement Values to be Output
A 3.5.3.1 GETOUTINFO_RS422, Request Data Selection
GETOUTINFO_RS422
The command lists all selected output data for the RS422 interface. The sequence shown corresponds to the
output sequence.
A 3.5.3.2 OUTADD_RS422, Selection of Data Additional Values
OUTADD_RS422 NONE| ([SHUTTER] [COUNTER] [TIMESTAMP] [INTENSITY] [STATE]
[DIST_RAW])
Selection of additional values to be transmitted.
- NONE: no output of additional values
- SHUTTER: output of exposure time
- COUNTER: output of measurement value counter
- TIMESTAMP: output of timestamp
- INTENSITY: output of intensity parallel to each distance value
- STATE: output of status word
- DIST_RAW: output of uncalibrated distance value (raw value)
A 3.6
Example Command Sequence During Selection of Measurement Value
Command
Content
MEASPEAK
Peak selection for displacement measurement
MEASRATE
Measuring rate (under consideration of reflectivity and movement
of the target)
OUTPUT
Selection of the output channel
OUTHOLD
Output characteristic during measurement errors
OUTADD_RS422
Selection of the additional values to be output for RS422
interface
BAUDRATE
Baud rate settings RS422 interface