RELAY AND CURRENT TESTING
d. With the STOP screen displayed, press <RIGHT> to tab along to the asterisk (*):
e. Press <ENTER> to stop the test process and switch the relay OFF.
7.2
Current Test
Its possible to test the current outputs following the procedure below:
1. Press <ENTER>: the ENTER PASSWORD screen is displayed.
2. Input your current security password; refer to section 4.5 Security Password.
The CALIBRATE menu is the first menu option displayed.
3. Press <RIGHT> four times until the TEST screen is displayed.
4. Press <ENTER> to access the TEST menu. From here you can access the Relay
and Current test options.
5. Set the Current test values to check the current output is reading correctly:
a. Press <RIGHT> until the TEST CURRENT screen is displayed.
b. Press <ENTER> to access edit mode.