XZR250 User’s Manual
Michell Instruments
9
INSTALLATION
Figure 9
Insert Sensing Module
1. Ensure housing is wired correctly; refer to section 2.2 Electrical Connections.
2. Hold the sensor head with the display facing downwards.
3. Fit sensing module to the sensor head and gently push to secure the module in
place; refer to Figure 9. The connector is keyed to aid alignment.
4. Keep the sensor head oriented with the sensor upwards until just before fitting the
probe to avoid dropping the sensor during installation.
5. With the clamp fitted over the sensor head shoulder carefully slide the head onto
the probe; refer to
Figure 10. Ensuring the probe is inserted fully.
6. Ensuring the display is correctly oriented, tighten the clamp: torque to 5Nm.
!
CAUTION: Take care not to bump or shock the probe and ensure
the gasket is not damaged when inserting the probe.
Figure 10
Fit sensor Sensor head to Probe Body