QMA401 User’s Manual
82
97450 Issue 2, March 2018
APPENDIX C
Register Configuration U — Internal Logging Configuration/Service Interval Days
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
RL
RL
RL
RL
DP
DP
DP
DP
DeltaF and ppm
V
Log Interval in cycles (DP)
Ram Buff er Log parameter (RL)
Range is 1 to 15 cycles. (for CAL use, def = 1)
0000 = Moisture – PPMv (def)
0001= Moisture - PPMw
0010 = Moisture – MGM3
0011 = Moisture – PA (wvp)
0100 = Moisture - LBMMSCF
0101 = Dew point
1111 = No logging
Register Configuration
V — Internal Logging Configuration/Service Interval Days
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
DC
DC
DC
DC
DC
DC
DC
DC
ML
ML
ML
ML
ML
ML
ML
ML
Dryer Capacity or Used (DC) – in ppm
Moisture Gen Capacity or Used (ML) – in
Days
0 to 255 x 100,000
represents 0 to 25,500,000 in 100,000 steps.
0 to 255 x 10
Represents 0 to 2,550 days (61200 hours) in 10
day steps
Register Configuration W — Signal Filter Settings
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
BF
BF
DF
DF
DF
DF
DF
BeatF Median Filter (BF) Setting
DeltaF Median Filter (DF) Sample size
5 sample Median fi lter to remove spikes
1=Median of 1 (middle value)
3=Median of 3, averaged (default)
Any other value = OFF
4-24 = Sample size of fi lter to smooth signal
(default=12)
< 4 or >24 = OFF
Summary of Contents for QMA401
Page 1: ...QMA401 Trace Moisture Analyzer User s Manual 97450 Issue 2 March 2018...
Page 71: ...QMA401 User s Manual Michell Instruments 61 APPENDIX A Appendix A Technical Specifications...
Page 75: ...QMA401 User s Manual Michell Instruments 65 APPENDIX C Appendix C Modbus Register Map...
Page 99: ......
Page 100: ...QMA401 User s Manual 90 97450 Issue 2 March 2018 NOTES...
Page 101: ...QMA401 User s Manual Michell Instruments 91 NOTES...
Page 102: ...http www michell com...