QMA401 User’s Manual
34
97450 Issue 2, March 2018
OPERATION
3.7.3 Outputs Screen
Figure 31
Outputs Screen
Parameter
Description
Output
Selector
Selects output required.
Available Options:
Output 1, Output 2
Parameter
Toggles through the different output parameters.
Available Options:
Oven ˚C, ml/min, Cell Pr. barg, External Pr. barg,
H
2
O ppm
V
, H
2
O ppm
W
, H
2
O mg/m
3
, WVP Pa, lbs/MMscf, DP ˚C
Output Type
Toggles the signal type of the output.
Available Options:
1-5 V, 4-20 mA
Min
Selects the lower output limit for the corresponding output.
Max
Selects the higher output limit for the corresponding output.
Error
Indication
Selects the mA error indication level for the outputs.
Available options:
3.2mA, 21.4mA
Table 6
Outputs Screen Parameters