Register Configuration E — Alarm Configuration
Note: Alarm 4 is a system fault/warning alarm and is configured in register 9
(Configuration M)
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
L4
L3
L2
L1
A3
A3
A3
A3
A2
A2
A2
A2
A1
A1
A1
A1
Alarm1 Parameter (A1)
Alarm2 Parameter (A2)
0000 = Moisture – PPMv (def)
0001= Moisture - PPMw
0010 = Moisture – MGM3
0011 = Moisture – PA (wvp)
0100 = Moisture - LBMMSCF
0101 = Dew point
0110 = Oven temperature
0111 = Flow rate
1000 = Cell Pressure
1001 = Ext Line Pressure
0000 = Moisture – PPMv
0001= Moisture - PPMw
0010 = Moisture – MGM3
0011 = Moisture – PA (wvp)
0100 = Moisture - LBMMSCF
0101 = Dew point
0110 = Oven temperature (def)
0111 = Flow rate
1000 = Cell Pressure
1001 = Ext Line Pressure
Alarm3 Parameter (A3)
Alarm Latch Control (L1 to L4)
0000 = Moisture – PPMv
0001= Moisture - PPMw
0010 = Moisture – MGM3
0011 = Moisture – PA (wvp)
0100 = Moisture - LBMMSCF
0101 = Dew point
0110 = Oven temperature
0111 = Flow rate (def)
1000 = Cell Pressure
1001 = Ext Line Pressure
L1 = 1 = Alarm1 latch
L2 = 1 = Alarm2 latch
L3 = 1 = Alarm3 latch
L4 = 1 = Alarm4 latch
L1 = 0 = Alarm1 don’t latch
L2 = 0 = Alarm2 don’t latch
L3 = 0 = Alarm3 don’t latch
L4 = 0 = Alarm4 don’t latch
Register Configuration F — Analog Output Config
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
T2
T1
O2
O2
O2
O2
O1
O1
O1
O1
Ouput1 Parameter (O1)
Output2 Parameter (O2)
Summary of Contents for QMA401
Page 1: ...QMA401 Trace Moisture Analyzer User s Manual 97450 Issue 2 March 2018...
Page 71: ...QMA401 User s Manual Michell Instruments 61 APPENDIX A Appendix A Technical Specifications...
Page 75: ...QMA401 User s Manual Michell Instruments 65 APPENDIX C Appendix C Modbus Register Map...
Page 99: ......
Page 100: ...QMA401 User s Manual 90 97450 Issue 2 March 2018 NOTES...
Page 101: ...QMA401 User s Manual Michell Instruments 91 NOTES...
Page 102: ...http www michell com...