QMA401 User’s Manual
Michell Instruments
7
INSTALLATION
2.2
Unpacking the Instrument
Open the crate and carefully unpack the instrument.
NOTE: Retain the packaging in case the instrument is returned for factory
calibration or service.
The accessories box should contain the following items:
•
Traceable calibration certificate
•
SD memory storage card
•
USB communications cable
•
IEC power cable
•
Application software CD
•
User's manual
If there are any shortages please notify the supplier immediately.
2.3
Electrical Requirements
The QMA401 requires the following electrical supply:
•
85 to 264 V AC,
47/63Hz, 250 VA max
•
Alarm outputs comprise four sets of changeover relay contacts, one set for
INTERNAL FAULTS, and three sets for PROCESS alarms. All contacts are rated
at 24 V, 1A.
NOTE: THIS RATING MUST NOT BE EXCEEDED
Fuse
This product is provided with an internally mounted fuse. A replacement fuse can be obtained
by contacting Michell Instruments' technical support. Fuse rating = 5 x 20mm 3 A anti-surge
to IEC 60127-2.
DANGER
Electric
Shock Risk
Power Connection
The product is provided with an IEC C13 socket on the rear panel for mains power input.
This equipment must be supplied with a voltage between the range of 85 to 264 V AC,
47/63Hz, maximum required power is 140 W.
Figure 3
Power Connection
Summary of Contents for QMA401
Page 1: ...QMA401 Trace Moisture Analyzer User s Manual 97450 Issue 2 March 2018...
Page 71: ...QMA401 User s Manual Michell Instruments 61 APPENDIX A Appendix A Technical Specifications...
Page 75: ...QMA401 User s Manual Michell Instruments 65 APPENDIX C Appendix C Modbus Register Map...
Page 99: ......
Page 100: ...QMA401 User s Manual 90 97450 Issue 2 March 2018 NOTES...
Page 101: ...QMA401 User s Manual Michell Instruments 91 NOTES...
Page 102: ...http www michell com...