17
Check the expiration date before performing a control
solution test. Do not use if expired. Please write down the
expiration date on the control solution bottle after first
opening.
Gently shake the control solution before use.
Discard the 1st drop of control solution, this will eliminate
any residue. Place a drop of control solution on a clean, dry
surface (e.g. the lid of the test strip vial).
Dip the test strip into the control solution. Results will
appear in 5 seconds. Compare the result to the range
printed on the test strip vial, the result should fall within
that range.
CAUTION
The control solution bottle icon contains the letter ‘c’ so it will not to be
confused with the medication bottle icon. This will allow you to differentiate
between a control solution test and an actual blood test.
NOTE
Repeat steps with both Low (Level 1) and High (Level 3) control solutions.