42
Message
What it means
Action required
Blood sample is placed
on the test strip before
the symbol for blood
blinks.
Wait for the meter to display the
blinking blood icon before applying
your blood sample.
1. Color area of the test
strip is dirty, defective
or there is too much
sunlight. Action
required.
2. Inserting another
brand of test strip in
the meter.
1. Please insert a new test strip and
perform your test again, or if it blinks
with “sun”, avoid direct sunlight and
retest. Contact customer support if
the problem persists.
2. Please insert the correct test strip.
The ambient
temperature is less
than 50˚F (10˚C).
Place the meter at a temperature
between 50 ~ 104˚F (10 ~ 40˚C) for
more than 10 minutes and test again.
The ambient
temperature is over
104˚F (40˚C).
Place the meter at a temperature
between 50 ~ 104˚F (10 ~ 40˚C) for
more than 10 minutes and test again.
The test result is less
than 20mg/dL.
If “Lo” appears on the LCD screen for
the Glucose result, the measured
concentration is below 20mg/
dL. The test should be repeated to
ensure that the test procedure was
done correctly. If you are certain the
meter is functioning correctly and
no errors were made during the test
procedure and the blood glucose
is still consistently measured as
“Lo”, the result may indicate severe
hypoglycemia (low blood glucose).
You should inform your Healthcare
Professional immediately.
Summary of Contents for EasyTouch BluLink
Page 1: ...Instructions for Use...
Page 47: ...47...