In the default range setting function mode, the maximum amplitude is
15V, and the minimum amplitude is 0.15V; the incremental step value is
0.01V (10mV).
The -20dB output attenuator may be selected by pressing the
button.
With attenuation, the output signal has a maximum value of 1.500V, the
minimum value of 0.015V, and the minimum step is 0.001V (1mV).
4.2.4
Setting CH1 voltage offset
Change the mode to bias adjustment, and move the cursor to the desired
digit, then turn the "ADJUST" knob to adjust the offset parameter.
4.2.5
Setting CH1 duty cycle
Change the mode to adjust the duty cycle, move the cursor to the desired
decimal digit, then turn the "ADJUST" knob to adjust the duty cycle pa-
rameter.
4.2.6
Adjusting phase difference between channels
Change the mode to adjust the phase, move the cursor to the desired decimal
position , then turn the "ADJUST" knob to adjust the phase parameter.
3
3
The phase difference only makes sense if CH1 and CH2 output the same frequency.
12