Scale Calibration Procedure (lbs. or kg) with Remote Display
12. Remove the remaining half load from the scale then press clear (C). The message
done
will flash briefly on the display.
Figure 48.
13.
Exit calibration mode.
Press and hold the test button for at least 3 seconds then release. With no load on the scale the
display should read 0.00 lb. or 0.000 kg.
Figure 49.
14. The unit’s calibration must now be verified as required by state and/or local Weights and Measures regulations (starting
on page 52).
Need to Start Over?
If for any reason you need exit the calibration mode or restart the process press the test button then the clear button.
Figure 50.