FDI2056 Use
r’s Manual
v 2.0 r 1.1
– 03/13
Copyright © 2013 Metrolab Technology SA
39
www.metrolab.com
-420
Query UNTERMINATED
The host is trying to read a response without having sent
a complete query.
-440
Query
UNTERMINATED
after indefinite response
Indicates that a query was received in the same program
message after a query requesting an indefinite response
was executed. On the FDI 2056, the only command
r
eturning an indefinite response (“Arbitrary ASCII
Response Data”) is *IDN.
100
INSTRUMENT-DEPENDENT COMMAND ERRORS
101
Invalid value in list
One or more values in a numeric list parameter are
invalid, e.g. floating point when not allowed.
102
Wrong units for parameter
A parameter within the command has the wrong type of
units for the command.
103
Invalid number of
dimensions in channel
The channel list provided refers to more layers of
multiplexers than allowed.
105
Numeric suffix invalid
The channel number provided is not valid.
200
INSTRUMENT-DEPENDENT EXECUTION ERRORS
200
Software Error
The firmware has encountered an unexpected error.
201
Data not all available
The user is trying to fetch more data than was acquired.
202
Data buffer was overrun
No probe plugged in.
203
Unit already exists
The user is trying to register a unit which already exists
in the system.
204
Unknown unit
The user is requesting information regarding an
inexistent unit.
205
Invalid encoder
configuration
The encoder configuration given is incorrect.
206
Operation in progress
The instrument is currently operating and cannot accept
the command.
207
Channels
don’t share the
same configuration
The command requested the value on the default
channel, but this value is not shared by all channels
present in the system.
208
Parameter value has been
rounded or modified to meet
internal constraints
The user has specified a value for a parameter that
cannot be completely satisfied. The value has been
rounded to the nearest value achievable by the system
or the value resulted in a change in the system
configuration.