:SENSe#:CORRection:SLOPe[:ACQuire]
(Event)
3-7-26
Proceed to an offset and slope correction on a channel. When proceeding to the
offset correction, the input is not shorted. It’s up to the user to proceed to that operation
if needed.
Syntax
SENSe#:CORRection:SLOPe[:ACQuire]
Example
SENS:CORR:SLOP
Starts an offset and a slope correction.
Related commands
INPut#:COUPling
SENSe#:CORRection:MEMorized
:SENSe#:SAMPler[:RATe]
3-7-27
Defines at which rate the sampler must acquire the incoming signal.
Syntax
SENSe#:SAMPler[:RATe] <value HZ> | DEFault
SENSe#:SAMPler[:RATe]? [DEFAult]
Default
500 KHZ
Example
SENS:SAMP 100 KHZ
Set the analog-to-digital converter sampling rate to 100 [kHz].
:SENSe#:FUNCTion
3-7-28
Defines how the sense subsystem must consider the incoming signal.
Syntax
SENSe#:FUNCTion
OFF|VOLTage|FLUX|DEFault
SENSe#:FUNCTion?
[DEFAult|OPTions]
Default
FLUX
Arguments
OFF
Turns the sense subsystem off.
VOLTage
Senses the input voltage.
FLUX
Integrates the input voltage.
Example
SENS:FUNC OFF