FDI2056 Use
r’s Manual
v 2.0 r 1.1
– 03/13
Copyright © 2013 Metrolab Technology SA
27
www.metrolab.com
Syntax
INPut#:GAIN
MINimum|MAXimum|UP
|DOWN|DEFault|<value>
INPut#:GAIN?
[DEFault|OPTions]
Default
0.1
Arguments
MINimum
Sets the input gain to its minimal value.
MAXimum
Sets the input gain to its maximal value.
UP
Increases the gain using the next valid value.
DOWN
Decreases the gain using the next valid value.
value
Valid values are 0.1, 0.2, 0.5, 1.0, 2.0, 4.0, 5.0, 10.0, 20.0,
40.0, 50.0, 100.0. Any other value will generate an error.
Example
INP:GAIN 10
Sets the gain of all channels to 10.
:MEASure#[:SCALar]:VOLTage?
(Query only)
3-7-20
Instantly measures a single non-averaged voltage value.
Syntax
MEASure#[:SCALar]:VOLTage? [<digits>]
Default
digits = 6
Arguments
digits
The number of digits that must be returned to the user.
Example
MEAS:VOLT? 3
This command might return
0.123 V \n
Related commands
FORMat#:DATa
FORMat#:READings:ALL
FORMat#:UNIT[:ENABle]
UNIT#:VOLTage