3-3 IEEE 488.2 / SCPI STATUS REGISTERS
IEEE 488.2 compliant instruments have at least two registers: the Status Byte and
the Standard Event Status Register. SCPI adds the Operation Status Register,
Questionable Status Register and Error/Event Queue. The diagram below, taken from
the SCPI standard, provides a good summary. This section describes how the
FDI 2056 uses these status registers.
Status Byte
3-3-1
Contains a 1-byte status summary. The FDI 2056 uses the following bits:
Bit
Name
Description
2
EAV
Error AVailable (in Error/Event Queue)
3
QSB
Questionable Summary Bit
4
MAV
Message AVailable: response ready to be read
5
ESB
Event Summary Bit
6
RQS
ReQuest for Service