![Metrohm Spectrosense Manual Download Page 40](http://html1.mh-extra.com/html/metrohm/spectrosense/spectrosense_manual_1779500040.webp)
5 Appendix
36
Spectrosense, Instructions for Use
•
Carry out adjustment
Carry out the voltage adjustment as described in
Titrando with
tiamo
Connect the Spectrosense to your Titrando (see
tiamo
software.
•
Enter Spectrosense into sensor list
If the Spectrosense has not yet been inserted into the sensor list you
can do this now in the program part
Configuration
under
Sensors
.
Click
Edit, New
and choose
Other sensor...
. Enter a
Sensor name
and choose the
Device
and the
Measuring input
. Confirm the entry
with
[OK]
.
•
Create method
Switch to the program part
Method
. With
File, New
you can create a
Blank method
and confirm your selection with
[OK]
.
With
Insert, New command
enter a new measuring command
MEAS U
and confirm with
[OK]
. Doubleclick the measuring com-
mand in order to open the corresponding properties window.
Adapt the following parameters in the tab
General/Hardware
:
Device
name
connected
Titrando
Measuring input
1 or 2
Sensor
previously entered Spectrosense
Stirrer
off
In the tab
Measuring parameters
change these parameters:
Measurement without drift control
activate
Measuring time
999'999
s
Measuring interval
0.5
s
In order to be able to start the method you first have to save it with
File, Save
. Choose the
Method group
in which you want to store the
method, enter a
Method name
and click
[Save]
.
•
Start measurement
Switch to the program part
Workplace
. In the sub window
Run
under
Sample data
select the method you previously created. The meas-
urement can now be started with
[Start]
.