2 Working with the Spectrosense
26
Spectrosense, Instructions for Use
2 Working with the Spectrosense
2.1 Preparing the Spectrosense
1 Attach sleeve
Insert the
electrode shaft
(
7
) into the
sleeve
(
6
) until its O-ring
snaps into the lower groove. This might be easier if you first moisten
the O-ring of the sleeve. For sample changer applications move the
sleeve to the upper groove. This way the electrode shaft protrudes
farther into the titration vessel.
2 Attach mirror
Screw the enclosed
mirror
(
8
) onto the end of the
electrode
shaft
(
7
).
3 Connect Spectrosense to titration device
Connect the
measuring output 'out'
(
2
) with the enclosed
elec-
trode cable 6.2116.020
with the measuring input
'Ind.'
of your
Titrino or Titrando.
4 Power supply
•
Power supply via Titrando
Connect the
power supply
(
1
) of the Spectrosense with an
MSB connection
of your Titrando. Use the enclosed
cable
6.2151.070
.
Spectrosense
2
1
6.2
151
.07
0
6.2
11
6.0
20
Fig. 2 Titrando – Spectrosense