5 Detailed description
774 Oven Sample Processor, Instructions for use
118
HTime is the additionally granted waiting time for still reaching the target
temperature. HAction is executed after HTime elapses.
The SCAN timeout is triggered on execution of a SCAN command. STime
determines the reliable response time of the connected device (generally
the determination time). SAction is executed after expiry of STime.
cont:
Continue without action
STOP: Stop
method
ERROR:
Trigger error message
5.11.1.13
M
ode
.G
as
.U
nitFlow
mL/min
, L/h
M
ode
.G
as
.M
inFlow
0
…999
M
ode
.G
as
.Ma
xFlow 0…
900
…999
M
ode
.G
as
.S
elect
Air
, others, N2
M
ode
.G
as
.O
therfac 0.001…
1
…9.999
.UnitFlow determines the unit of the gas flow rate. The two nodes .Minflow
and .Maxflow define the limits which, if undershot or overshot, trigger error
messages (RS error E163 resp. E206).
Node .Otherfac takes effect only if 'others' is selected at the .Select node.
5.11.1.14
M
ode
.D
osimatSet
.D
osUnitNo
1
…12
M
ode
.D
osimatSet
.1.DosR
ate 0.01…160
ml/min,
max.
M
ode
.D
osimatSet
.1.F
illRate 0.01…160
ml/min,
max.
M
ode
.D
osimatSet
.1.DosT
ube
1
...4
M
ode
.D
osimatSet
.1.FillT
ube 1...
2
...4
M
ode
.D
osimatSet
.1.Ex
chTube 1...
2
...4
M
ode
.D
osimatSet
.1.P
repTube
1
...4
M
ode
.D
osimatSet
.1.Em
ptyTube 1...
4
etc.
until
.12
Dosing unit settings.
DosUnitNo:
Number of current dosing unit
DosRate: Dosing
speed
FillRate: Filling
speed
DosTube:
Dosing outlet of 700 Dosino
FillTube:
Filling inlet of 700 Dosino
ExchTube:
Rinsing inlet on exchanging the exchange unit (see
&Assembly.Dos.Value("release") 5.11.6.6)
PrepTube:
Dosing outlet on preparing cycle of 700 Dosino
EmptyTube:
Air inlet for emptying the 700 Dosino.
5.11.1.15
M
ode
.M
an
S
top
.R
emCtl
STOP device1
, 14 Bit (1,0, or
∗
)
M
ode
.M
anStop
.RS
Ctl
&M;$S,
14 ASCII characters
M
ode
.M
anStop
.F
low
cont.
, off
M
ode
.M
anStop
.H
EATER
INIT
, off, 50…250
Functions executed when the <STOP> key is pressed.
.RemCtl
Bit pattern defines the line statuses of the Remote socket
.RSCtl
Character string is transferred via the RS232 cable
.Flow
The pump and the solenoid valve remain in the
current state or are both deactivated
.HEATER
Oven control is either set to the initial temperature
or switched off. Any temperature may also be
selected