5 Detailed description
774 Oven Sample Processor, Instructions for use
104
Detailed Status Conditions
Status conditions of the global $R:
$R.Mode
Basic state: ready to start automatic processing
$R.Assembly
An assembly step has been executed.
Status conditions of the global $G:
$G.Mode.Start.
Instrument at the beginning of processing
$G.Mode.Start.01.WAIT
Instrument processing the start sequence, displays line number and
current command
$G.Mode.Sample.01.WAIT
Instrument processing the sample sequence, displays line number and
current command
$G.Mode.Final.01.WAIT
Instrument processing the final sequence, displays line number and
current command
$G.Mode.
Instrument processing a manual command
$G.Assembly.
Instrument processing an assembly-command
Status conditions of the global $H:
$H.Mode
The status conditions of the global $H are identical with the ones of the
global $G.
5.9.5 Error Messages, Errors
Fatal Instrument Errors:
E1
Incorrect program check sum
E2
RAM read/write error
E3
RAM lost data
E4
Timer interrupt for multi-tasking missing
E5
RS232 module test error
E6
RS232 read/write error
E7
Display read/write error
E12
EBUS
error
E18
Low battery
E19
RAM test error
Program-specific Messages:
E26
Manual stop
E28
Wrong object call
E29
Wrong value or no value allowed
E30
Wrong
trigger
E31
Comman currently not allowed
RS Receive Errors:
E36
Parity
error
Exit: <QUIT> and set the same parity for both instruments.
E37
Stop bit error
Exit: <QUIT>and set the same stop bit for both instruments.
E38
Overrun error. At least 1 character could not be read.
Exit: <QUIT>.
E39
Internal receive buffer full (>82 characters).
Exit: <QUIT>.
Summary of Contents for 774