6.4
Diagnosis
774 Oven Sample Processor, Instructions for use
145
6.4.3 Main memory (RAM)
This diagnosis step completes a non-destructive test of the entire area of
the RAM contents (main memory).
•
Prepare the unit for diagnosis (see
•
If necessary, press <
Ð
> several times until
diagnosis
>RAM test
•
<ENTER>
If no errors are found, you will see the following on the display :
>RAM test
RAM test ok
•
<ENTER>
6.4.4 Display
This diagnosis step allows you to check the LEDs and the display for oper-
ability.
•
Prepare the unit for diagnosis (see
•
If necessary, press <
Ð
>several times until
diagnosis
>display test
•
<ENTER>
After you press key <ENTER>, the program runs automatically through a test run for optical
checking of the LEDs and the display.
⇒
All LEDs blink consecutively for a brief period.
⇒
The background lighting of the display is switched off for a short period and back on again.
⇒
The power-up test pattern (each pixel active) is displayed.
⇒
Both lines of the display are cleared.
⇒
Both lines of the display consecutively display characters "
#
", "
H
" and, lastly, "
I
".
⇒
Both lines display the continuous moving text
"0123456789ABCDEFGHIJKLMNOPQRSTUVWXYZ" from right to left.
•
The test run can be halted and started again by pressing key <5>.
•
The test is quit with key <QUIT> or <STOP>.
diagnosis
>display contrast test
•
<ENTER>
After you press the <ENTER> key, the following display is shown, whereby the contrast of the
display is varied constantly between bright and dark.
>display contrast test
774 Oven Sample Proc.
•
The test can be quit with key <QUIT> or <STOP>.
Summary of Contents for 774