5 Detailed description
774 Oven Sample Processor, Instructions for use
128
5.11.6 &Assembly …
5.11.6.1
A
ssembly
.S
ample $G
A
ssembly
.S
ample
.F
unc
=
, +, –
A
ssembly
.S
ample
.V
alue 0...
1
...999
Defines the (first) sample beaker (rack position) to be processed. Modifica-
tion of the sample variable.
.Func Function
.Value
Value (absolute or relative)
&Assembly.Sample;$G triggers this function.
5.11.6.2
A
ssembly
.M
ove $G,
$S
A
ssembly
.M
ove
.T
arget 1,
2
A
ssembly
.M
ove
.P
osition
sample
, spec.1…8, 0...999
Positioning a beaker at the specified tower.
.Target
Target or tower
.Position
Rack position or identification of a specific vial
&Assembly.Move;$G triggers this function.
5.11.6.3
A
ssembly
.L
ift $G,
$S
A
ssembly
.L
ift
.S
tation
1
, 2,
∗
A
ssembly
.L
ift
.W
ay
rest
, work, rinse, shift, special, 0...325 mm
Move lift.
.Station
Selection of lift station (
∗
= both lifts)
.Way
absolute lift position
&Assembly.Lift;$G triggers this function.
5.11.6.4
A
ssembly
.H
eater $G
A
ssembly
.H
eater.
T
emp
Init
, off, 50…250
A
ssembly
.H
eater
.Ti
me
blank
, 1...999
Temperature control.
.Temp
Target temperature in °C
.Time
Heating interval in minutes
&Assembly.Heater;$G triggers this function.
5.11.6.5
A
ssembly
.F
low $G
A
ssembly
.F
low
.A
ddress
Pump
, Valve
A
ssembly
.F
low
.S
tate
on
, off
Pump or valve control.
.Address
Selection of pump or valve
.Value State
&Assembly.Flow;$G triggers this function.
5.11.6.6
A
ssembly
.D
os $G,
$S
A
ssembly
.D
os
.A
ddress
1
...12
A
ssembly
.D
os
.V
alue
±0.01…999.999 ml, fill, release,
prepar., empty , eject, adjust, level
Control of dosing units.
.Address
Selection of dosing unit
.Value
Volume or function
&Assembly.Dos;$G triggers this function.