4 Installation
■■■■■■■■
17
a) Assembling the electrode holder
CAUTION
Before assembling the electrode holder make sure that all parts are
completely dry. Check especially
■
the connection socket for the SPE
■
the slot in the silicone seal
for solution. Liquid entering the connection socket for the SPE
damage the electrode holder.
1
Place the silicone seal
in the shaft of the SPE holder
. The
support the positioning of the seal in the shaft and
prevent it from turning out of position.
2
Place the supporting ring
on top of the seal with the planar side
pointing towards the seal. The two pins
ing. The slot in the supporting ring has to be congruent with the slot
in the silicone seal.
3
Place the nut
on the shaft. Do not tighten the nut, otherwise
the SPE cannot be assembled with the holder.
b) Assembling the SPE
CAUTION
Always hold screen-printed electrodes at the edges, in order not to
touch the active electrode surface.
1
Align the screen-printed electrode with the symbol on the shaft of
the electrode holder. The symbol and the electrode surface have to
point in the same direction.