Retest period
Retest period in months.
Options:
Editable
– allows Retest period to be modified while running Auto
Sequence®. Numeric keypad for entering custom Retest period is
offered within the test.
Not editable
– Default Retest period is used. Retest period cannot be
modified while running Auto Sequence®.
Note
This flow command is active only if Auto Sequence
®
is started from the Auto Sequences
®
Main menu.
Inspection Expert mode
If Inspection Expert mode flow command is set, the Visual inspection screen and Functional
inspection screen within Auto Sequence
®
are displayed for 1 second and an overall PASS is
automatically applied at the end of test. In between, the automatic procedure can be stopped
and statuses can be applied manually.
Inspection Expert mode is disabled by default.
Parameters
State
On
– enables automatic settings of tickers in Visual and Functional tests.
Off
– disables automatic settings of tickers in Visual and Functional tests.
Flow Protocol
This flow command controls commands for communication with external device for the control
of the flow of Auto Sequence
®
.
Parameters
Communication
settings
Selects the port for communication with the external device.
-
RS232(PC)
-
USB
Flow settings
Commands for communication with external device
(industrial PC for example)
Proceed
Send string:
Proceed
Description: If Proceed Flow setting is enabled,
implementation of Auto Sequence
®
on instrument can be
automatically controlled by external device by using
Proceed
string command. Command works in parallel
with OK / ENTER / TEST / HV TEST keys and the
EXTERNAL TEST / OK key mode. Proceed Flow setting is
disabled by default.