MI 3394 CE
MultiTester
XA
Single tests
106
(1) peak voltage
(4) Ulim
(2) voltage at disconnection time
(5) moment of disconnection
(3) calculated voltage value
(6) discharging time
Figure 6.62: Measuring principle (external)
Test circuit (Output = External)
Figure 6.63: Discharging Time test (Output = External)
Discharging Time test procedure (Output = External)
Select the
Discharging Time
function.
Set test parameters / limits.
Connect test leads to the DISCHARGING TIME terminals on the instrument and on the
device under test.
Connect device under test to the mains supply and Switch it ON.
Start measurement.
Measurement is stopped manually by disconnecting device under test mains supply.
Save results (optional).