MI 2014 Cable Scanner
Section I General information
5
Section I General information
2. Safety and operational precautions
2.1 Warnings
Ø
Never connect the test equipment to an active network.
Ø
Service is allowed to be carried out only by an authorised person!
Ø
Use only standard or optional test accessories supplied by Metrel!
Ø
Use only connector types equivalent to those built in to avoid damage to the
instrument components.
Ø
If the test equipment is used in a manner not specified in this Users Manual,
the protection provided by the equipment may be impaired!
2.2 Battery replacement
Note
•
Insert cells correctly, otherwise the instrument will not operate and the batteries
could be discharged.
•
If the instrument is not to be used for a long period of time, remove all batteries from
the battery compartment.
•
Battery compartment is protected with a fuse to prevent abnormal use of battery.
•
External charger is not intended for supplying the instrument without batteries.
Warnings
Ø
Do not charge alkaline batteries.
Ø
Use only chargers delivered from Metrel or distributor of the test equipment to
avoid possible fire or electric shock.
Ø
In case of blown fuse consult your distributor.
2.3 Service and recalibration
It is essential that test instrument is regularly calibrated in order technical specification
listed in this Instruction Manual to be guaranted. We recommend the calibration to be
carried out once per 2 years.
Metrel encloses to every new instrument an original calibration certificate.
For recalibration and repairs under or out of warranty time please contact your
distributor for further information.