MI 3110 EurotestIM
Technical specifications
52
Reference temperature range ........... 10
C
30
C
Reference humidity range ................. 40 %RH
70 %RH
Operation conditions
Working temperature range .............. 0
C
40
C
Maximum relative humidity ............... 95 %RH (0
C
40
C), non-condensing
Storage conditions
Temperature range ........................... -10
C
+70
C
Maximum relative humidity ............... 90 %RH (-10
C
+40
C)
80 %RH (40
C
60
C)
Communication transfer speed ........ RS 232 57600 baud
............................................ USB
256000 baud
The error in operating conditions could be at most the error for reference conditions
(specified in the manual for each function) +1 % of measured value + 1 digit, unless
otherwise specified in the manual for particular function.