A 1636 

DC/AC Current clamp  

 

5  Technical Specifications 

 

Function 

Measurement  of  AC/DC  current,  phase,  power 
energy, power, power factor 

Power supply 

Zn-Mn dry cell 6F22 9VDC (external power source is 
allowed) 

Rated current 

15mA 

Test mode 

Clamp CT 

Clamp Size 

Φ

68mm 

Range 

DC 0.0A~2000A; AC 0.0A~1000A 

Resolution 

0.1A AC/DC  

Signal output 

1 mV/1A 

Accuracy 

±2.0%FS  (50Hz/60Hz;  23

±2

,  below  70%RH, 

keep the wire be in the center of clamp) 

Phase Error 

≤3°(AC50Hz/60Hz; 23

±2

 

Zero clearing 

Press  zero  key  is  to  clear  the  data,  eliminate  the 
effect  of  geomagnetic  field  to  the  measurement  of 
AC current 

Output connection 

Standard  connection:  Red  wire-positive  output; 
Yellow wire-negative output.  

Output Wire Length 

2m 

Measured Wire 

Position 

Approximately  in  the geometric center of the clamp, 
the  errors  for  deviating  the  center  of  the  clamp 
increase 1.5%rdg.  

Current Frequency 

AC:45Hz-400Hz 

Voltage of circuit 

Below AC 600V 

Voltage of battery 

When  the  voltage  of  battery  declines  to  7.2V,  and 
the  LED  light  will  be  on  to  indicate  replace  the 
battery. 

Dimension 

224mmx115mmx43mm 

Weight 

515g 

Working 

Environment 

-10

~45

below 80%rh 

Storage 

Environment 

-10

~60

; below 70%rh

 

Insulation Strength  

AC 3700V/rms (between coil and shell) 

Safety Rules 

IEC1010-1,  IEC1010-2-032,  Pollution  degree:2  CAT 

(600V) 

 

Summary of Contents for A 1636

Page 1: ...DC AC Current clamp 2000A A 1636 Instruction manual Version 1 2 Code no 20 752 824...

Page 2: ...rel metrel si 2012 METREL Mark on your equipment certifies that this equipment meets the requirements of the EU European Union concerning safety and electromagnetic compatibility regulations No part o...

Page 3: ...Current clamp 3 Table of contents 1 SYMBOLS AND WARNINGS 4 2 DESCRIPTION OF CURRENT CLAMP 5 3 OPERATION INSTRUCTIONS 6 4 MAINTENANCE 7 INSPECTION 7 CLEANING 7 SERVICE AND CALIBRATION 7 5 TECHNICAL SPE...

Page 4: ...rally known precautions in order to avoid risk of electric shock while dealing with electric installations Do not extend hands over safety barrier to prevent of electric shock Only handles are allowed...

Page 5: ...gn no need to disconnect the measured circuits non contact safe and fast It can be connected with phase detection analyzer industrial control equipment data recorder oscilloscope harmonic analyzer ele...

Page 6: ...and the output of sensor is 1mV 1A it means that when the input current is 1A the ratio metric voltage is1mV 1 Clamp 2 Up cover 3 Trigger Open and close the clamp 4 Indicate of positive input AC curr...

Page 7: ...ave them to dry totally before using it Do not use liquids based on petrol or hydrocarbons Do not spill cleaning liquid over the current clamps To clean jaw cut surfaces use slightly oiled soft cloth...

Page 8: ...magnetic field to the measurement of AC current Output connection Standard connection Red wire positive output Yellow wire negative output Output Wire Length 2m Measured Wire Position Approximately in...

Reviews: