A 1322 / A 1422
Technical specifications
36
8 Technical
specifications
8.1 Test functions
8.1.1 Earth bond resistance
Refer to chapter
Technical specifications
, paragraph
Earth bond resistance
in
master instrument user manual for detailed information.
8.1.2 Insulation
resistance (including S-probe)
Refer to chapter
Technical specifications
, paragraphs
Insulation resistance
and
Insulation resistance S-probe
in master instrument user manual for detailed
information.
8.1.3 Substitute leakage current (including S-probe)
Refer to chapter
Technical specifications
, paragraphs
Substitute leakage
current
and
Substitute leakage current S-probe
in master instrument user
manual for detailed information.
8.1.4 Differential leakage current
Differential leakage current readout
Range Resolution
Accuracy
0.00 mA
9.99 mA
0.01 mA
(5 % of r 5 digits)
Frequency response complies to EN61010 – Figure A1.
Pass levels
*
........................... 0.25 mA, 0.50 mA, 0.75 mA, 1.00 mA, 1.50 mA, 2.25 mA,
2.50 mA, 3.50 mA, 4.00 mA, 4.50 mA, 5.00 mA, 5.50 mA,
6.00 mA, 7.00 mA, 8.00 mA, 9.00 mA
Test duration* ....................... 2 s, 3 s, 5 s, 10 s, 30 s, 60 s, 120 s, 180 s, none
Test terminals ....................... Test socket (A 1322 / A 1422: 16A-5p, 32A-5p or 16A-3p)
Additional error ..................... 0.01 mA / A
*
Pass level and test duration are set on master instrument.
8.1.5 Touch leakage current
Refer to chapter
Technical specifications
, paragraphs
Touch leakage current
in master instrument user manual for detailed information.