4.5 Messages
When master instrument works with A1322/A1422 some new messages are used.
Refer to user manual of master instrument for detailed information.
Message Description
SigmaGT is not connected
to A1322 (A1422).
No connection between PAT mains test socket and A
1322 / A 1422 PAT test socket (5) !
Check connection / cable.
Connect PAT and
A1322/A1422 via RS232.
No serial communication (RS232) connection between
PAT and A1322/A 1422 . Check serial communication
cable.
MEASUREMENT
ABORTED
COMMUNICATION
INTERRUPTED.
Serial communication between master instrument and
A1322/A1422 was lost.
MEASUREMENT
ABORTED BUCAUSE OF
SAFETY!
Instrument skipped the required test because of a failed
previous test.
MEASUREMENT
ABORTED.
Use 1-phase supply on
A1322/A1422.
For performing selected test/measurement 1-phase
supply voltage system is required.
For testing of RCDs with
Idn=100mA or 30mA,
A1322/1422 must be used.
Selected test/measurement can only be performed using
A1322/A1422. Connect A1322/A1422 to master
instrument and start test again.
For this measurement use
A1422.
Selected test/measurement can only be performed using
A1422. Connect A1322/A1422 to master instrument and
start test again.
EXTERNAL VOLTAGE ON
TEST SOCKET TOO HIGH
ON A1322/A1422.
DANGER!
-
Voltage on test socket (16A or 32A) is higher
than approximately 25 V (AC or DC)!
Disconnect the test probe from the DUT and determine
why external voltage was detected!
EXTERNAL VOLTAGE ON
TEST V16A/32A TOO
HIGH ON A1322/A1422.
DANGER!
-
Voltage on test plug (16A or 32A) is higher
than approximately 25 V (AC or DC)!
Disconnect the test probe from the DUT and determine
why external voltage was detected!
HARDWARE ERROR
Return the instrument to the
repair centre.
The instrument detects a serious failure.