A 1322 / A 1422
Measurements
20
5.4 Differential leakage current measurement
Refer to chapter
Measurements – Single tests for appliances
, paragraph
Differential leakage current
in master instrument user manual for detailed
information.
Leakage current is measured via power supply cord of tested appliance.
Procedure:
Item Comment
Prepare test system.
See
chapter 4.3 Working
with the adapter.
Connect power supply plug of tested appliance into
proper test socket.
See
figure below.
Start measurement on the master instrument.
Wait until measurement is finished or stop the
measurement.
Check result(s).
Store result(s).
Optional
Figure 5.4:
Differential leakage current measurement
Warning:
During the test, DUT is powered as for normal operation. Pay attention on
possible electric shock, and danger of rotating/moving parts!
Note:
Consider any displayed warning before starting measurement!